1586A
Calibration Manual
42
CALibrate:REFerence?
Description: Return nominal value expected for reference entry
Example:
CAL:REF?
Response: 3.00
Related
Commands:
CALibrate:NEXT
CALibrate:STATe?
CALibrate:INFOrmation?
CALibrate:STEP?
CALibrate:SECure:STATe <boolean>, <admin_password>
Description: Instruct Product to enable calibration
Example:
CAL:SEC:STATE OFF,”1586”
Related Commands:
CALibrate:SECure:STATe?
CALibrate:SECure:STATe?
Description: Query calibration enable state
Example:
CAL:SEC:STAT?
Response: 1
Related Commands:
CALibrate:SECure:STATe
CALibrate:SECTion
Description: Skip to next section of calibration procedure.
Example:
CAL:SECT
Related Commands:
CALibrate:BACKup
CALibrate:SKIP
CALibrate:SKIP
Description: Skip to next entry point in calibration procedure.
Example:
CAL:SKIP
Related Commands:
CALibrate:BACKup
CALibrate:SECTion