SUPER-DAQ Precision Temperature Scanner
Calibration Adjustment
29
Unlock the Product with a Remote Interface
To unlock the Product with a remote interface, send the command:
CAL:SEC:STAT OFF,“1586”
To relock the Product, send the command:
CAL:SEC:STAT ON
Reset the Admin Password
If the admin password is lost or forgotten, the password can be reset to 1586 with
these steps:
Note
Before doing these steps, try to use the factory default password:
1586
.
1. Do the general disassembly steps in the “Disassembly” section.
2. Connect a jumper across J8, as shown in Figure 8.
3. Reconnect the mains power cable between the Product and a power outlet.
4. Turn the on the Product.
5. When the Product is started, the password will automatically be reset to
1586
.
6. Turn the Product off and disconnect the mains power cable.
7. Remove the jumper connected in step 2.
8. Reassemble the Product by doing the reverse procedure in “Disassembly”.