Fluke Calibration 1586A Super-DAQ Calibration Manual Download Page 5

 

 iii 

List of Tables 

Table Title Page 

1. Symbols 

.....................................................................................................   2 

2. Fuses 

.........................................................................................................   14

 

3.  

Memory Clear Functions ...........................................................................   17

 

4. 

Required Test Equipment ..........................................................................   18

 

5. 

DC Volts Verification Steps .......................................................................   22

 

6. 

4-Wire Ohms Verification Steps ................................................................   23

 

7. 

PRT Verification Steps ..............................................................................   24

 

8. Thermistor 

Verification 

Steps 

.................................................................... 

 

25

 

9. DC 

Current 

Verification 

Steps 

................................................................... 

 

27

 

10. Adjustment 

Steps 

...................................................................................... 

 

33

 

11.  List of Commands .....................................................................................   37

 

12.  Full Calibration Example ...........................................................................   38

 

13.  1 V DC Calibration Example ......................................................................   39

 

14.  Write Calibration Date to a Module Example ............................................   39

 

15.  User-Replaceable Parts and Accessories .................................................   45

 

 

 

Summary of Contents for 1586A Super-DAQ

Page 1: ...luke Corporation All rights reserved Specifications are subject to change without notice All product names are trademarks of their respective companies 1586A SUPER DAQ Precision Temperature Scanner Ca...

Page 2: ...nearest Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB D...

Page 3: ...le Voltage Accuracy 9 Thermocouple Reference Junction Accuracy 9 Thermocouple Temperature Accuracy 10 Thermocouple Measurement Characteristics 11 DC Voltage 11 DC Voltage Accuracy 11 DC Voltage Input...

Page 4: ...e Accuracy Verification 27 1586 2588 Module CJC Accuracy Verification 28 Calibration Adjustment 28 Unlock the Product 28 Unlock the Product with a Remote Interface 29 Reset the Admin Password 29 Disas...

Page 5: ...ps 22 6 4 Wire Ohms Verification Steps 23 7 PRT Verification Steps 24 8 Thermistor Verification Steps 25 9 DC Current Verification Steps 27 10 Adjustment Steps 33 11 List of Commands 37 12 Full Calibr...

Page 6: ...1586A Calibration Manual iv...

Page 7: ...Battery Location 16 3 DC Voltage Test Connections 21 4 4 Wire Ohms Test Equipment Setup 23 5 PRT Test Equipment Setup 24 6 Thermistor Test Equipment Setup 25 7 mA DC Current Equipment Setup 26 8 Passw...

Page 8: ...1586A Calibration Manual vi...

Page 9: ...cept Contact Fluke Calibration To contact Fluke Calibration call one of the following telephone numbers Technical Support USA 1 877 355 3225 Calibration Repair USA 1 877 355 3225 Canada 1 800 36 FLUKE...

Page 10: ...voltage MAINS installation CAT III 1 Measurement Category III is applicable to test and measuring circuits connected to the distribution part of the building s low voltage MAINS installation CAT IV 1...

Page 11: ...e the mains power cord if the insulation is damaged or if the insulation shows signs of wear Make sure the ground conductor in the mains power cord is connected to a protective earth ground Disruption...

Page 12: ...annels to be hazardous live and an electric shock hazard if any channel is connected to a hazardous voltage source Do not remove touch or change the internal wiring of hazardous inputs until the input...

Page 13: ...agnetic Environment IEC 61326 1 Basic Controlled EM for full specification Radio Frequency Emissions IEC CISPR 11 Group 1 Class A Group 1 has intentionally generated and or uses conductively coupled r...

Page 14: ...unless otherwise noted after a warm up time of 1 hour and within an environment temperature range of 18 C to 28 C and may depend on the channel The confidence level for accuracy specifications is 95...

Page 15: ...nominal PRT RTD When using 3 wire PRT RTD add 0 039 C to the accuracy specification for internal resistance mismatch and voltage offset if using Channel 1 or add 0 15 C if using channels x01 through x...

Page 16: ...en using 2 wire thermistor add the number given in the table to the specification for internal resistance If the environment temperature is outside the specified range increase the accuracy specificat...

Page 17: ...0 0001 C 0 001 C 10 A 2 1 k to 98 k 0 0001 C 0 001 C 10 A 95 k to 1 M 0 0001 C 0 001 C 1 A Thermocouple Temperature Range 270 C to 2315 C depending on the sensor Voltage Range 15 mV to 100 mV Thermoc...

Page 18: ...0 56 C 1 14 C 0 68 C 0 55 C 0 57 C 1 26 C 0 76 C 0 60 C 0 63 C J 210 C to 1200 C 200 C 0 C 1000 C 0 20 C 0 08 C 0 11 C 0 29 C 0 12 C 0 14 C 0 65 C 0 28 C 0 25 C 1 41 C 0 61 C 0 53 C N 270 C to 1300 C...

Page 19: ...erature is outside the specified range multiply the temperature coefficient numbers by the temperature deviation and add to the accuracy specification Range Accuracy Channel 1 Ch x01 x20 Fast Sample R...

Page 20: ...dium or slow sample rate For 2 wire resistance add 0 02 internal resistance if using Channel 1 or 1 5 if using channels x01 through x20 and add external lead wire resistance When using Fast sample rat...

Page 21: ...bounce Time 1 7 ms Maximum Count 1048575 20 bits Trigger Absolute Voltage Range 4 V to 30 V Minimum Logic High 2 0 V Maximum Logic Low 0 7 V Minimum Pulse Width 50 s Maximum Latency 100 ms Alarm Outpu...

Page 22: ...ake sure that the mains power cord is firmly plugged into the power module on the rear of the Product 3 Make sure the power source that the Product is plugged into is energized 4 Make sure that the li...

Page 23: ...e power entry module 3 Insert a small flat screwdriver blade into the narrow recess to the left of the fuse holder and pry to the right until the holder pops out The Product is shipped with a replacem...

Page 24: ...ed for Product memory To replace the battery see Figure 2 1 Disassemble the Product as described in the Disassembly section 2 Remove the battery and replace with one rated appropriately for the select...

Page 25: ...iles Factory Reset Deletes test setup files DMM data files and scan data files from the internal memory 1 Clears the configuration of the channel setup test setup and instrument setup 2 1 Does not rem...

Page 26: ...e short or equivalent Fluke PN 2653346 Alternate standard The standard resistors which have equivalent spec with Fluke 5720A PRT Standard Fluke 5522A for 1 k value Standard resistors of 25 100 200 and...

Page 27: ...5 C 1586 2588 DAQ STAQ CJC Metrology Drywell Fluke 9171 with insert block Reference thermistor probe Fluke 5610 9 Accuracy 0 013 C or better at 25 C Reference thermometer Fluke 1586A Only for CJC veri...

Page 28: ...his section provides performance tests to verify that the Product operates within published specifications as well as a complete calibration adjustment procedure The performance test and if necessary...

Page 29: ...the keypad diagnostic program DC Volts Verification Connect the Product to the test equipment as shown in Figure 3 and apply the voltages listed in Table 5 Note For the 0 V tests use the 4 wire short...

Page 30: ...2 mV 99 9928 mV 0 04 1 0 1 39 995 mV 40 005 mV 0 1 1 0 1 99 9928 mV 100 0072 mV 0 1 7 0 V 7 0 V 1 1 1 1 000032 V 0 999968 V 1 1 1 0 999968 V 1 000032 V 0 10 50 0 V 50 0 V 10 1 10 10 00029 V 9 99971 V...

Page 31: ...WIDEBAND NC hcn205 eps Figure 4 4 Wire Ohms Test Equipment Setup Table 6 4 Wire Ohms Verification Steps Nominal Input Range 1 Year Test Limits High Low 0 100 0 0035 0 0035 100 100 100 0075 99 9925 0...

Page 32: ...table use standard resistors Configure Ch001 as a 4 wire PRT function with resistance display 5522A Calibrator 1 ke UUT NC NC hcn312 eps Figure 5 PRT Test Equipment Setup Table 7 PRT Verification Ste...

Page 33: ...the resistance listed in Table 8 Configure Ch001 as a 4 wire thermistor function with resistance display 5522A Calibrator UUT NC NC hcn204 eps Figure 6 Thermistor Test Equipment Setup Table 8 Thermis...

Page 34: ...Hi source to Lo Product and Lo source to mA Product For the 0 A tests open Lo inputs 5520A CALIBR ATOR HI LO TRIG GUAR D TC 20A NORMAL AUX SCOPE OUT V RTD A SENSE AUX V 20V PK MAX 20V PK MAX FUSED X 1...

Page 35: ...be used with 8508A to obtain suitable test uncertainty ratio 1586 2586 CJC Module Accuracy Verification 1 Connect the calibrated E type thermocouples to channel 10 on the Product UUT 2 Insert the ther...

Page 36: ...vals A one year interval is adequate for most applications Product accuracy specifications are not valid beyond the one year interval Adjustments are accessed through both the remote interface and fro...

Page 37: ...ith these steps Note Before doing these steps try to use the factory default password 1586 1 Do the general disassembly steps in the Disassembly section 2 Connect a jumper across J8 as shown in Figure...

Page 38: ...30 J8 RST PSW Fan Battery Display Cable InGuard Shield A6 PCB Assembly Short J8 then turn on the power After the instrument is successfully started the admin password is reset to 1586 hcn103 eps Figur...

Page 39: ...all High Capacity Modules and test leads from the Product 2 Turn off the mains power at the rear of the Product and remove the mains power cord The front panel power key only puts the Product into a p...

Page 40: ...1586A Calibration Manual 32 To Remove Pull End From Unit 3 Boot Removal 1 Battery 4 5 2 6 hcn208 eps Figure 9 Disassembly...

Page 41: ...ustment Product value or range that is to be adjusted Amplitude of the adjustment signal The frequency of the adjustment signal if necessary Table 10 Adjustment Steps Step Value Range Input Signal Des...

Page 42: ...0 k N R_190_2 190 0 N R_0_2 0 0 N R10K 10 0E 3 Start calibration of FRES 10 k range Y R_10K_3 10 0 k N R_1_9K_3 1 9 k N R_0_3 0 0 N R100K 100 0E 3 Start calibration of FRES 100 k range Y R_100K_4 100...

Page 43: ...N PRT P400FR 400 0 Start calibration of PRT 400 range Y P_400_FR 400 0 N P_200_FR 200 0 N P_0_FR 0 0 N THERM T2K 2 2E 3 Start calibration of FTH 2 k range Y T_1_9K_1 1 9 k N T_1_0K_1 1 0 k N T_190_1...

Page 44: ...1 100 0 A N DC_0A_1 Open N DC_N100UA_1 100 0 A N DC1MA 1 0E 3 Start calibration of dc 1 mA range Y DC_P1_0MA_2 1 0 mA N DC_0A_2 Open N DC_N1_0MA_2 1 0 mA N DC10MA 10 0E 3 Start calibration of dc 10 mA...

Page 45: ...ge or instructions associated with running step CALibrate MOD DATE slot year month day Set CAL date of module in specified slot CALibrate NEXT reference Continue a calibration procedure if it is waiti...

Page 46: ...calibrator to the instrument for DCV calibration CAL NEXT Show instruction to ask for 100 mV signal input and the reference value CAL NEXT 0 1 Continue to perform measurements for this point after it...

Page 47: ...s for this point after it is completed show instruction to ask for 1 V signal input and the reference value CAL NEXT 1 Continue to perform measurements for this point after it is completed calculate t...

Page 48: ...ation procedure Example CAL BACK Related Commands CALibrate SKIP CALibrate SECTion CALibrate CONSt cco_name Description Retrieves the value in use of the given calibration constant Example CAL CONS DC...

Page 49: ...t Description Set and query module calibration date Example CAL MOD DATE 1 2013 11 1 CAL MOD DATE 1 Response 2013 11 1 Related Commands CALibrate SECure STATe CALibrate NEXT reference Description Cont...

Page 50: ...able calibration Example CAL SEC STATE OFF 1586 Related Commands CALibrate SECure STATe CALibrate SECure STATe Description Query calibration enable state Example CAL SEC STAT Response 1 Related Comman...

Page 51: ...e should be disabled Example CAL_START MAIN CAL_START MAIN DC1V Related Commands CALibrate SECure STATe CALibrate STATe Description Return state of calibration Example CAL STAT Response RUN running a...

Page 52: ...Manual 44 CALibrate STORe CALibrate STORe Description Store new calibration constants or query whether a cal store is needed Example CAL STOR CAL STOR Response 1 is yes 0 if no Related Commands CALib...

Page 53: ...IN FLAT PHILLIPS SS PASSIVATED LOCK PATCH ROHS COMPLIANT Contact Fluke 1 FLUKE 1586A 2586 HI CAPACITY MODULE Not Shown 1588940 1 BATTERY PRIMARY LI MNO2 3V 560MAH COIN CR2450 24X5MM BULK Not Shown 434...

Page 54: ...1586A Calibration Manual 46 1 2 3 4 6 5 7 8 hcn209 eps Figure 10 Replaceable Parts...