SUPER-DAQ Precision Temperature Scanner
Performance Tests
25
Thermistor Verification Steps
Connect the Product to the test equipment as shown in Figure 6 and apply the
resistance listed in Table 8. Configure Ch001 as a 4-wire thermistor function with
resistance display.
5522A Calibrator
UUT
NC
NC
hcn204.eps
Figure 6. Thermistor Test Equipment Setup
Table 8. Thermistor Verification Steps
Nominal
Input
Range
1-Year Test Limits
High Low
0
Ω
2.2
k
0.2
Ω
-0.2
Ω
2 k
Ω
2.2
k
2000.28
Ω
1999.72
Ω
90 k
Ω
98
k
90004.1
Ω
89995.9
Ω
900 k
Ω
1
M 900140
Ω
899860
Ω