Victoreen 8000
Operators Manual
4-14
4.6 Waveforms - Partial kV Waveforms
Exposure Results, Low Sensitivity
kVp Avg
kV Eff
kV Peak
mSec
mR
40.5 39.7 42 10.6
2
kV Waveform
Low Sensitivity
0
5
10
15
20
25
30
35
40
45
0
2
4
6
8
10
12
mSec
kV
kV
NOTE
In the above kV waveform, no data is plotted below
approximately 39 kV. This is because there is
insufficient data available to fully construct the kV
waveform. This occurs when exposures are made
near the detectability limit of the NERO mAx. To
remedy this situation, switch to high sensitivity or to
the next lower filter range. Below is the kV
waveform from the same exposure made using
high sensitivity.
Exposure Results, High Sensitivity
kVp Avg
kV Eff
kV Peak
mSec
mR