Victoreen 8000
Operators Manual
2-38
9.8
R/min
HVL INTERN
RATE LOW
MODE CHAMBER
MODE
SENS
SAVE
After the exposure has started, the NERO mAx displays the exposure rate. When the rate has stabilized
or the desired rate has been achieved, press the SAVE softkey (softkey 4) to save the rate measurement.
9.8
R/min
HVL INTERN
RATE LOW
TURN OFF
MODE CHAMBER
MODE
SENS
BEAM
Turn off the x-ray beam. The NERO mAx has stored the measured exposure rate and will set up for the
next exposure as soon as the beam turns off.
0.0
R/min
Please Wait…
HVL INTERN
RATE LOW
MODE CHAMBER
MODE
SENS
REJECT
While the exposure rate from the first exposure is displayed, the NERO mAx prepares to take the second
exposure. Please wait while the NERO mAx prepares to take the next exposure. If the NERO mAx
detects a fault condition such as a high offset, an error message is displayed and the user must correct
the error to continue (see Section 4.3--Error Messages--for more information).
9.8
R/min
MAKE EXPOSURE # 02
HVL INTERN
RATE LOW
MODE CHAMBER
MODE
SENS
REJECT
The NERO mAx is now ready to take the next exposure. Position an aluminum absorber in the beam and
make the exposure. If the previously measured exposure is not acceptable, press the reject softkey to
retake the exposure. If an overrange occurs, an error message is displayed and the exposure must be
taken over.
Summary of Contents for Victoreen NERO mAx 8000
Page 62: ...Victoreen 8000 Operators Manual Blank page...
Page 84: ...Victoreen 8000 Operators Manual Blank page...
Page 88: ...Victoreen 8000 Operators Manual Blank page...
Page 92: ...Victoreen 8000 Operators Manual Blank page...
Page 94: ...Victoreen 8000 Operators Manual Blank page...
Page 97: ...Blank page...