4
Troubleshooting
Waveforms-Partial kV Waveforms
4-15
In the
exposure
below, part of the kV waveform appears to disappear after an initial peak. This is because after the initial
peak, the radiation output drops below the level at which the NERO mAx can accurately reproduce the kV
waveform. This level is approximately 400 counts on the channel A radiation waveform.