![Fluke Biomedical Victoreen NERO mAx 8000 User Manual Download Page 85](http://html1.mh-extra.com/html/fluke-biomedical/victoreen-nero-max-8000/victoreen-nero-max-8000_user-manual_2310746085.webp)
B
Appendix
mA Limits vs. kV
B-1
Appendix B
mA Limits vs. kV
B.1 mA Limits vs. kV
mA Limits vs. kV at 26" SDD, 1.2 mm Al Inherent Filtration - Low Sensitivity:
W/Al, 30 - 60 kV Filter
Low Sensitivity
0
100
200
300
400
500
600
1
2
3
4
kV
mA
30
40
50
60
mA Limits vs. kV at 26" SDD, 1.2mm Al Inherent Filtration - High Sensitivity:
W/Al, 30 - 60 kV Filter
High Sensitivity
0
200
400
600
1
2
3
4
kV
mA
30
40
50
60