![FlowLine LI50-1001 Instruction Manual Download Page 39](http://html1.mh-extra.com/html/flowline/li50-1001/li50-1001_instruction-manual_2308219039.webp)
LI5X Process & Temperature Meter Instruction Manual
39
Low Alarm with Fail-Safe Operation (Set < Reset)
Reset
Set
Input
Relay
LED
Automatic (non-latching)
de-energized
energized
on
off
Relay
LED
Automatic or Manual (non-latching)
ACK
pressed
Relay
LED
Manual (latching)
ACK
pressed
Relay
LED
Manual only after passing above Reset (latching with clear)
ACK
pressed
Fail-safe operation:
relay coil is energized in non-alarm condition. In
case of power failure, relay will go to alarm state.
Spec Tech Industrial Electric
www.spectechind.com
888-773-2832
Spec Tech Industrial Electric
www.spectechind.com
888-773-2832