Configuration Using Display
Step Four
EchoPod
®
can be configured before installation. The transmitter features non-volatile memory, so any setting
configured before installation will not be lost when the switch is powered down. To configure, follow the steps
below:
1. Measure the Tank
a. Begin by measuring the key tank and fitting dimensions.
b. Correct tank dimensions will result in accurate sensor measurement.
2. Set the Units of Measurement (
UNITS
)
a. Units can be configured in basic engineering units of length including Feet, Inches, Centimeters
or Meters.
3. Set the Sensor Height (
HEIGHT
)
a. This is the empty setting (4mA) for the tank.
4. Set the Fill-Height (
FILL-H
)
a. This is the full setting (20mA) for the tank.
5. Set the Fail-Safety (
SAFE
)
a. This is the maximum measurement range for the sensor. The sensor will ignore all echo returns
beyond this setting.
NOTE:
When configuring EchoPod
®
, choose either the WebCal
®
or Push Button method. Either method will
accomplish the goal of sensor configuration. Changes to the configuration can be made using the alternative method.
When beginning with one method, it is recommended to complete the configuration before using the other method to
make any adjustments.