![FlowLine EchoPod UG01-03 Manual Download Page 15](http://html1.mh-extra.com/html/flowline/echopod-ug01-03/echopod-ug01-03_manual_2308254015.webp)
MN310140 Rev A2
15
|
Configuration Using WebCal
®
Step Three
EchoPod
®
can be configured before installation. The transmitter features non-volatile memory, so any setting
configured before installation will not be lost when the switch is powered down. To configure, follow the steps
below:
1. Install WebCal
®
software
a. Go to
www.flowline.com/webCal-software/
and select language version.
b. Review how USB
®
Fob interfaces with EchoPod
®
and your computer.
2. Measure the Tank
a. Begin by measuring the key tank and fitting dimensions.
b. Correct tank dimensions will result in accurate sensor measurement.
3. Sensor Configuration
a. Configures Loop Fail-Safe, Output at Empty & Startup Condition for the sensor.
4. Dimensional Entry
a. Distance Mode (default)
i. Basic information for operation (Sensor Height & Fill-Height).
5. Tank Level Confirmation
a. Confirm the values are accurate for the application.
6. Write to Unit
a. Uploads configuration into the sensor.
b. Access to a customer wiring diagram specific to the relay configuration.
Note:
When configuring EchoPod
®
, choose either the WebCal
®
or Push Button method. Either
method will accomplish the goal of sensor configuration. Changes to the configuration can be
made using the alternative method. When beginning with one method, it is recommended to
complete the configuration before using the other method to make any adjustments.