Maintenance
7.1 Cleaning and Storage
Wipe the housing with a damp cloth as needed. Please do not use abrasives or solvents
to clean the meter. If the meter is not to be used for an extended period, remove the
batteries and store them separately.
7.2 Battery and Fuse Replacement
WARNING: To avoid electrical shock, disconnect the meter from any connected circuits,
remove the test leads from the meter terminals, and set the function switch to the OFF
position before attempting to replace the batteries or fuses.
1.
To replace the batteries, unlock the battery compartment
by rotating the lock with a screwdriver or other tool (the
lock is item 1 in the accompanying diagram).
2.
Remove the battery compartment cover.
3.
Replace the two (2) 1.5V ‘AAA’ batteries, observing correct
polarity.
4.
Replace the battery cover and re-lock it before operating
the meter.
5.
To replace the fuses, lift the tilt stand and locate the two
screw holes (item 2 in the diagram).
6.
The screws are protected by rubber grommets which must
be removed before the screws can be accessed.
7.
Remove the two screws and remove one additional screw
located just behind the battery compartment lock to
access the fuse area.
8.
Replace Fuse F1 for
A/mA current input with a 0.4A/600V DC/AC, IR 30kA F fuse or
better; Dimension: 6 x 32 mm
9.
Replace Fuse F2 for ‘A’ current input with a 11A/600V DC/AC, IR 20kA F fuse or
better; Dimension: 10 x 38mm
10.
Secure the meter before operating.
7.3 Disposal of Electronic Waste
As with most electronic products, this equipment must be disposed of in an
environmentally friendly way, and in accordance with existing regulations for electronic
waste. Please contact your FLIR Systems representative for more details.