FLIR DM6x USER MANUAL Document Identifier: DM6x-en-US_AA
26
VFD_ACV (with Low Pass Filter) DM64/DM66 only
RANGE
Accuracy
1)
10Hz ~ 100Hz (fundamental)
600.0V
1.0% + 3d
100Hz ~ 400Hz (fundamental)
600.0V
10% + 3d
2)
1)
Not specified for fundamental frequency > 400Hz
2)
Accuracy linearly decreases from 1% + 3d @100Hz to 10% + 3d @400Hz
Auto Voltage ACV (DM64/DM66)
RANGE
Accuracy
1)
45Hz ~ 440Hz
600.0V
2.0% + 3d
1)
Not specified at <1VAC
Threshold: > 1VAC nominal
Approximate input impedance (//164pF) for reference:
At direct input
50Vac (typical) from quiescence:
>8M
@ < 5.6Vac
22k
@ 7Vac
12k
@ 8Vac
2.6k
@ 50Vac
At direct input >>50V (typical) from quiescence:
Initial impedance is approximately 2.1k
. Impedance increases abruptly within a
fraction of a second as display voltage (hard signal) is much higher than 50V (typical).
Final impedances vs display voltages typically are:
12k
@100V
100k
@300V
240k
@600V
DC Voltage
RANGE
Accuracy
60.00mV, 600.0mV, 6.000V
0.3% + 2d
60.00V
0.4% + 2d
600.0V
0.2% + 2d
Input Impedance: 10M
, 54pF nominal