![Flintec MCS-64 Manual Download Page 38](http://html1.mh-extra.com/html/flintec/mcs-64/mcs-64_manual_2305544038.webp)
Page Manual MSC-64
38
ST
Set Tare
RPDO1 [ 00 08 ]
This command will activate the net weighing function by storing the current weight value as a
tare.
The weight signal must be “stable” within the limits set by NR (No Motion Range) and NT (No
Motion Time) commands for the “signal stable” bit to be active and set tare command to be
accepted.
Master (PC / PLC) sends
Device responds
Result
ST OK
Tare performed
If the weight signal is “stable”, the response to the IS command (Device Status) will show the
“signal stable” bit active and the ST command will be accepted (OK). If the “signal stable” bit
is not active, the ST command will be rejected and the LDU will respond with ERR (error).
RT
Reset Tare
RPDO1 [ 00 04 ]
The weighing signal returns to gross mode.
Master (PC / PLC) sends
Device responds
Result
RT OK
Tare deactivated
The LDM88.x responds to the RT command with either OK or ERR. If OK is returned then
the “tare active” bit in the Device Status (IS) response will be set to “0”.