Zero tracking will be performed only on results less than +/-0.5 d at a rate of 0.4 d/sec, where
d = display step size (see DS command). The zero can only be tracked to +/- 2% of
maximum (see CM command).
In non legal for trade applications you can exceed the limit +/- 2%. If you require for example
a zero tracking range of +/- 10 d then you have to set ZT to be 100 where d is the division
size (which means d * 10).
Factory default: ZT=0
FD
Factory default settings
[ 2006sub02 ]
This command puts the LDM88.x back to a known state. The data will be written to the
EEPROM and the TAC will be incremented by 1.
Note:
All calibration and setup information will be lost by issuing this command!
Master (PC / PLC) sends
Device responds
Result
CE
E+00017
(example)
Request: TAC-counter CE 17
CE_17 OK
Calibration sequence active
FD OK
Factory default setting
CS
Save the calibration values
[ 2004sub02 ]
This command results in the calibration values being saved to EEPROM, and causes the
TAC to be incremented by 1.
Master (PC / PLC) sends
Device responds
Result
CE
E+00017
(example)
Request: TAC-counter CE 17
CE_17 OK
Calibration sequence active
CS OK
Calibration values saved
The CS command saves all of the calibration group values, as set by CZ, CG, CM, DS, DP
and ZT. The command returns ERR and has no updating action unless it is preceded by the
CE_XXXXX.