Page Manual MSC-64
32
AZ
Absolute zero point calibration
[ 2300sub02 ]
The command AZ is used as reference point for all weight calculations and will setup in
mV/V. Permitted values are ± 32000 (= ± 3.2000 mV/V).
Master (PC / PLC) sends
Device responds
Result
AZ
Z+0.0005
Request :
Zero point @ 0.0005mV/V
CE
E+00017
(example)
Request: TAC-counter CE 17
CE_17 OK
Calibration sequence active
AZ_00500 OK
Setting:
new: Zero point @ 0.0500 mV/V
Factory default: 00000 @ 0.0000 mV/V input signal.
AG
Absolute gain calibration
[ 2300sub01 ]
The command AG is used as absolute gain (or measuring range) for all weight
calculations and will setup in mV/V. Permitted values are ± 32000 (= ± 3.2000 mV/V).
Master (PC / PLC) sends
Device responds
Result
AG
G+2.0000
Request :
Meas. range 2.000 mV/V
CE
E+00017
(example)
Request: TAC-counter CE 17
CE_17 OK
Calibration sequence active
AG_19998_90900
OK
new: Measuring range
1.9998 mV/V @ 90900 d
Factory default: 200000 @ 2.0000 mV/V input signal.
If the AG value has to be set at 2.0000 mV/V, but the load cell capacity is quoted at
2.2 mV/V, you will have to ratio the load cell capacity down.
Example : A 100kg load cell with an Output at Rated Load of 2.2 mV/V would be equivalent
to 90.9 kg @ 2 mV/V (exact 1.9998 mV/V)
ZT
Zero
tracking
[ 2100sub12 ]
This command enables or disables the zero tracking. Parameter = 0 disables the zero
tracking and parameter = 1 enables the zero tracking. Issuing the command without any
parameter returns the current ZT value.
Master (PC / PLC) sends
Device responds
Result
ZT Z:001
Request : ZT status (ON)
CE
E+00017
(example)
Request: TAC-counter CE 17
CE_17 OK
Calibration sequence active
ZT_0 OK
Setting: ZT = OFF