FD
Factory default settings
This command puts the DAS 72.1 back to a known state. The data will be written to the EEPROM and the
TAC will be incremented by 1.
Note:
All calibration and setup information will be lost by issuing this command!
Master ( PC / PLC ) sends
DAS 72.1 responds
Result
CE
¿
E+00017
(example)
Request: TAC-counter CE17
CE_17
¿
OK
Calibration sequence active
FD
¿
OK
Factory default setting
CS
Save the calibration values
This command results in the calibration values being saved to EEPROM, and causes the TAC to be
incremented by 1.
Master ( PC / PLC ) sends
DAS 72.1 responds
Result
CE
¿
E+00017
(example)
Request: TAC-counter CE17
CE_17
¿
OK
Calibration sequence active
CS
¿
OK
Calibration values saved
The CS command saves all of the calibration group values, as set by CZ, CG, CM, DS, DP and ZT. The
command returns ERR and has no updating action unless it is preceded by the CE_XXXXX.
Page 36