C.4.2 Set up Commands – SD, MT, GA, TE, TR, TL, AI
Note:
All setups should be stored with the WP command before power off.
SD
Start Delay 0 … 500 ms
Set the delay between falling/rising edge of trigger pulse and start of measurement. Permitted values are 0
… 500 ms.
Master ( PC / PLC ) sends
DAS 72.1 responds
Result
SD
¿
S+00100
Request: SD=100 ms
SD_200
¿
OK
Setting: SD=200 ms
Factory default: 0 [= 0 ms]
See time diagram of check weighing page 32
MT
Measuring Time, range 0…500 milliseconds
Set the time over which the average value will be built. Permitted values are 0 … 500 ms.
Master ( PC / PLC ) sends
DAS 72.1 responds
Result
MT
¿
M+00100
Request: MT=100 ms
MT_500
¿
OK
Setting: MT=500 ms
Note:
MT = 0 means disabled trigger and average function.
Factory default: 0
See time diagram of check weighing page 32
GA
Get Average
Issuing the GA command the DAS 72.1 returns the current average weight value by using the MT setup.
Master ( PC / PLC ) sends
DAS 72.1 responds
Result
GA
¿
A+01100
Answer: GA=1100 g
Note:
During the time between the trigger condition being accepted and the average value being updated,
the GA command will return the value 99 999.
TE
Trigger Edge
Issuing the TE command selects rising or falling edge trigger. Parameter = 0 select falling edge and
parameter = 1 select rising edge.
Master ( PC / PLC ) sends
DAS 72.1 responds
Result
TE
¿
E:001
Request: TE=1
TE_0
¿
OK
Setting: TE=0
Factory default: 0 [falling edge]
See time diagram of check weighing page 32
Manual DAS 72.1
Page 31