DS
Display step size
This command allows the output to step up or down by a unit other than 1. Permitted values are 1, 2, 5, 10,
20, 50, 100 and 200.
Master ( PC / PLC ) sends
DAS 72.1 responds
Result
DS
¿
S+00002
Request: display step size 2
CE
¿
E+00017
(example)
Request: TAC-counter CE17
CE_17
¿
OK
Calibration sequence active
DS_50
¿
OK
Setting: step size 50
DP
Set decimal point position
This command allows the decimal point to be positioned anywhere between leftmost and rightmost digits of
the 5-digit output result. Position 0 means no decimal point.
Master ( PC / PLC ) sends
DAS 72.1 responds
Result
DP
¿
P+00002
Request: position of decimal point
CE
¿
E+00017
(example)
Request: TAC-counter CE17
CE_17
¿
OK
Calibration sequence active
DP_0
¿
OK
Setting: no decimal point
CZ
Set calibration zero point
This is the reference point for all weight calculations, and is subject to TAC control.
Master ( PC / PLC ) sends
DAS 72.1 responds
Result
CE
¿
E+00017
(example)
Request: TAC-counter CE17
CE_17
¿
OK
Calibration sequence active
CZ
¿
OK
Zero point saved
Factory default: approx. 0 mV/V input signal
CG
Set calibration gain (span) value
This is the reference point for calibration under load, and is subject to TAC control.
Permitted values are 1 … 99 999.
Master ( PC / PLC ) sends
DAS 72.1 responds
Result
CG
¿
G+10000
Request: measuring range 10 000 d
CE
¿
E+00017
(example)
Request: TAC-counter CE17
CE_17
¿
OK
Calibration sequence active
CG_15000
¿
OK
Setting: span 15 000 d
For calibration an input signal near the display maximum (CM) will give the best system performance. The
minimum calibration load of at least 20% is recommended.
Factory default: 20 000 = 2,000 mV/V input signal
Page 34
Manual DAS 72.1