![Flexim FLUXUS WD100 User Manual Download Page 87](http://html1.mh-extra.com/html/flexim/fluxus-wd100/fluxus-wd100_user-manual_2304737087.webp)
A Menu Structure
UMFLUXUS_F5WDV1-1EN, 2016-04-08
87
A
Menu Structure
INIT-
resistant
Program Branch
Parameter
main menu: selection of the program branch
Parameter
input of the outer pipe diameter
input of the pipe circumference
This display will only be indicated if
Special Funct.\SYSTEM set-
tings\Dialogs/Menus\Pipe Circumfer.
is activated and
Outer Diam-
eter
=
0
has been entered.
input of the pipe wall thickness
range: depends on the connected transducers
default: 3 mm
selection of the pipe material
input of the sound speed of the pipe material
range: 600...6553.5 m/s
This display will only be indicated if
Other Material
has been selected.
selection whether the pipe is lined
selection of the lining material
This display will only be indicated if
Lining
=
yes
has been selected.
input of the sound speed of the lining material
range: 600...6553.5 m/s
This display will only be indicated if
Other Material
has been selected.
input of the liner thickness
default: 3 mm
input of the roughness of the inner pipe wall
range: 0...5 mm
default: 0.1 mm (for steel as pipe material)
input of the fluid temperature
default: 20 °C
input of the length of an extension cable
>PAR< mea opt sf
Parameter
Outer Diameter
100.0
mm
Pipe Circumfer.
314.2
mm
Wall Thickness
3.0
mm
Pipe Material
↕
Carbon Steel
c-Material
3230.0
m/s
Lining
no
>YES<
Lining
↕
Bitumen
c-Material
3200.0
m/s
Liner Thickness
3.0
mm
Roughness
0.4
mm
Medium Temperat.
20.0
C
Additional cable
65.0
m
Summary of Contents for FLUXUS WD100
Page 1: ...User Manual UMFLUXUS_F5WDV1 1EN Ultrasonic Flowmeter for Water FLUXUS WD100 FLUXUS WD200...
Page 4: ...FLUXUS WD100 WD200 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Page 8: ...FLUXUS WD100 WD200 Table of Contents 8 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Page 104: ...FLUXUS WD100 WD200 D Declarations of Conformity 104 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Page 105: ......
Page 106: ......