11 Advanced Measuring Functions
UMFLUXUS_F5WDV1-1EN, 2016-04-08
53
Select the command
→
Clear totalizer
in the upper line to reset the totalizers to zero. Press ENTER.
Automatic toggling between the displays
Select
Special Funct.\SYSTEM settings\Storing\Toggle totalizer
to toggle automatically between the
flow totalizer displays for the positive and negative flow direction.
Selection of the totalizers for storing
It is possible to store only the value of the totalizer that is currently displayed or one value for each flow direction. Select
Special Funct.\SYSTEM settings\Storing\Quantity Storage
.
When the measurement is stopped
The behavior of the totalizers when the measurement is stopped or after a RESET of the transmitter is set in
Special
Funct.\SYSTEM settings\Measuring\Quantity recall
.
11.3.1
Overflow of the Totalizers
The overflow behavior of the totalizers can be set:
Without overflow:
• The value of the totalizer increases to the internal limit of 10
38
.
• If necessary, the values will be displayed as exponential numbers (±1.00000E10). The totalizer can only be reset to zero
manually.
With overflow:
• The totalizer will be reset to zero automatically when ±9999999999 is reached.
Select
Special Funct.\SYSTEM settings\Measuring\Quant. wrapping
.
Independently of the setting, the totalizers can be reset to zero manually.
This error message will be displayed if the totalizers of a measuring channel used for mea-
suring the flow velocity are to be activated. The flow velocity cannot be totalized.
Enter a time interval between 0 (off) and 5 s. Press ENTER.
If
one
is selected, only the totalizer whose value is changing will be stored. This can apply
to the totalizer for the positive or the negative flow direction.
If
both
is selected, the values of the totalizers for both flow directions will be stored.
Press ENTER.
If
on
is selected, the values of the totalizers will be stored and used for the next measure-
ment.
If
off
is selected, the totalizers will be reset to zero.
Select
on
to work with overflow. Select
off
to work without overflow. Press ENTER.
Note!
The transmission of the sum of both totalizers (the throughput
Q
) via an output will not be valid after
the first overflow (wrapping) of one of the corresponding totalizers.
To signalize the overflow of a totalizer, an alarm output with the switching condition
QUANT.
and the
type
HOLD
have to be activated.
NO COUNTING
!
3.5
m/s
Toggle totalizer
0
s
Quantity Storage
one
>BOTH<
Quantity recall
off
>ON<
Quant. wrapping
off
>ON<
Summary of Contents for FLUXUS WD100
Page 1: ...User Manual UMFLUXUS_F5WDV1 1EN Ultrasonic Flowmeter for Water FLUXUS WD100 FLUXUS WD200...
Page 4: ...FLUXUS WD100 WD200 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Page 8: ...FLUXUS WD100 WD200 Table of Contents 8 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Page 104: ...FLUXUS WD100 WD200 D Declarations of Conformity 104 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Page 105: ......
Page 106: ......