Supplement to User Manual - ENGLISH
66
8
Request of Additional Quantities
8.1
Channel-Independent Diagnostic Values
The absolute and relative difference values of measuring channel A and B are channel-independent diagnostic values.
They are available in blocks of 16 Modbus registers in a predefined address range.
The start address is 30000.
For the calculation of the Modbus register address, an offset is added to the start address of the channel.
For the calculation of the Modbus register address, an additional offset is added to the start address (see Tab. 6.3).
Determination of the State
The state depends on the state of the values of measuring channel A and B:
• If the state of the value of measuring channel A and/or B is "not available" or "failed", the channel-independent diagnostic
value has the same state.
• If the state of the value of both measuring channels is "ok", the state of the channel-independent diagnostic value is also
"ok" and the time since the last update is reset.
8.2
Additional Physical Quantities and HPI Quantities
Tab. 8.1: Offset of the channel-independent diagnostic values
value
offset
predefined unit of measurement
channel-independent diagnostic values
flow velocity - absolute
0
dependent on the value of measuring channel A
flow velocity - relative
16
%
sound speed - absolute
32
dependent on the value of measuring channel A
sound speed - relative
48
%
signal amplitude - absolute
64
dependent on the value of measuring channel A
signal amplitude - relative
80
%
SNR - absolute
96
dependent on the value of measuring channel A
SNR - relative
112
%
SCNR - absolute
128
dependent on the value of measuring channel A
SCNR - relative
144
%
Example:
diagnostic value: sound speed - absolute
parameter: state
Modbus register address: 30000 + 32 + 0 = 30032
Tab. 8.2: Offset HPI
value
offset
unit of mea-
surement
data format
state of measuring channel
2
600
integer 16 bit
time since last update
601
s
integer 16 bit
volumetric flow rate
positive flow direction
602
m³/h
floating point number 32 bit
negative flow direction
604
floating point number 32 bit
totalizer for positive flow direction
606
m³
integer 32 bit
totalizer for negative flow direction
608
integer 32 bit
state
3
610
integer 16 bit
*
*
Summary of Contents for FLUXUS ADM 5 07 Series
Page 4: ...Modbus Server Supplement to User Manual 4 SU_ModbusV2 8 2015 03 02...
Page 6: ...Modbus Server Erg nzung zur Bedienungsanleitung DEUTSCH 6 SU_ModbusV2 8 2015 03 02...
Page 46: ...Modbus Server Erg nzung zur Bedienungsanleitung DEUTSCH 46 SU_ModbusV2 8 2015 03 02...
Page 48: ...Modbus Server Supplement to User Manual ENGLISH 48 SU_ModbusV2 8 2015 03 02...