Supplement to User Manual - ENGLISH
61
7
Request of Multiple Measured Values or Diagnostic Values
With one request, it is possible to provide
• multiple scaled measured values or
• the state of multiple scaled measured values or
• multiple diagnostic values or
• the state of multiple diagnostic values
of a channel.
The Modbus registers are predefined in a continuous address range.
For requesting data from Modbus registers of non-continuous address ranges see annex B.
Request of Multiple Measured Values
For the start addresses and offsets see Tab. 7.1 and Tab. 7.2.
The measured values are references to the scaled hi and lo word (see Tab. 6.3). They are displayed with the unit of mea-
surement defined by the user.
Tab. 7.1: Start addresses of the channels
channel
start address
A
15000
B
16000
C
17000
D
18000
Z
19000
Y
20000
Tab. 7.2: Offset of the measured value
physical quantity (floating point number 32 bit)
offset
lo word
hi word
medium temperature T
fluid
(at the location where the flow rate is measured, supply line or return
line)
0
1
medium temperature T
aux
(the other temperature, return line or supply line)
2
3
medium pressure p
fluid
(at the location where the flow rate is measured, supply line or return line) 4
5
medium pressure p
aux
(the other pressure, return line or supply line)
6
7
signal amplitude
8
9
sound speed
10
11
flow velocity
12
13
volumetric flow rate
14
15
volumetric flow rate, totalizer for positive flow direction
16
17
volumetric flow rate, totalizer for negative flow direction
18
19
standard volumetric flow rate (gas measurement)
20
21
standard volumetric flow rate, totalizer for positive flow direction
22
23
standard volumetric flow rate, totalizer for negative flow direction
24
25
mass flow rate
26
27
mass flow rate, totalizer for positive flow direction
28
29
mass flow rate, totalizer for negative flow direction
30
31
heat flow rate
32
33
Summary of Contents for FLUXUS ADM 5 07 Series
Page 4: ...Modbus Server Supplement to User Manual 4 SU_ModbusV2 8 2015 03 02...
Page 6: ...Modbus Server Erg nzung zur Bedienungsanleitung DEUTSCH 6 SU_ModbusV2 8 2015 03 02...
Page 46: ...Modbus Server Erg nzung zur Bedienungsanleitung DEUTSCH 46 SU_ModbusV2 8 2015 03 02...
Page 48: ...Modbus Server Supplement to User Manual ENGLISH 48 SU_ModbusV2 8 2015 03 02...