Settings in the Transmitter
2.1
Input of the HotCode
FLUXUS ADM 7x07, F705, G70x, PIOX TS37x, S705
• Press key C.
• Enter HotCode
485000
.
FLUXUS ADM 5x07, ADM 8x27, ADM 8x27P, F808, G80x, G80xP
2.2
Selection of the Mode
2.3
Input of the Transmission Parameters
Select
Special Funct.\SYSTEM settings\Miscellaneous
.
Select
yes
to enter a HotCode.
Enter HotCode
485000
. Press ENTER.
Select the mode.
•
sender
: The transmitter is operated as sender via the RS485 interface.
•
Modbus
: The transmitter is operated as Modbus slave.
Press ENTER.
Select
Special Funct.\SYSTEM settings\Network
to change the settings of the
transmission parameters.
Enter the device address.
Range: 1...247
Every device connected to the Modbus has to have an individual device address.
Select
default
to display the preset transmission parameters.
Select
setup
to change the transmission parameters.
Press ENTER.
Set the transmission parameters:
Select a list item in the upper line.
Change the setting in the lower line. Press ENTER.
•
baud
: baud rate (9 600)
•
parity
:
parity (
EVEN
)
•
st
: number of stop bits (1)
The number of data bits cannot be changed.
The transmission parameters can only be changed if
setup
has been selected.
SYSTEM settings
Miscellaneous
Input a HOTCODE
no
>YES<
Please input a
HOTCODE: 000000
RS485 interface
sender
>MODBUS<
SYSTEM settings
↕
Network
Device address:
2
ADR
RS485 protocol
default
>SETUP<
>BAUD< parity
st
9600 EVEN
1
Summary of Contents for FLUXUS ADM 5 07 Series
Page 4: ...Modbus Server Supplement to User Manual 4 SU_ModbusV2 8 2015 03 02...
Page 6: ...Modbus Server Erg nzung zur Bedienungsanleitung DEUTSCH 6 SU_ModbusV2 8 2015 03 02...
Page 46: ...Modbus Server Erg nzung zur Bedienungsanleitung DEUTSCH 46 SU_ModbusV2 8 2015 03 02...
Page 48: ...Modbus Server Supplement to User Manual ENGLISH 48 SU_ModbusV2 8 2015 03 02...