27
IEC 61000-4-11
<5 % UT
(>95% dip in UT.)
for 1 cycle
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95 % dip in UT)
for 250 cycles
<5 % UT
(>95% dip in UT.)
for 1 cycle
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95 % dip in UT)
for 250 cycles
interruptions, it is recommended that
the models Rooter
®
X3000 be powered
from an uninterruptible power supply
or a battery.
Power frequency
(50/60 Hz)
magnetic field
lEC 61000-4-8
30A/m
30A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Table 3: Guidance & Declaration - electromagnetic immunity concerning Conducted RF &
Radiated RF
The model Rooter
®
X3000 is intended for use in the electromagnetic environment specified below. The customer
or the user of the models Rooter
®
X3000 should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment - guidance
Conducted RF lEC
61000-4-6
Conducted RF lEC
61000-4-6
Radiated RF lEC
61000-4-3
3 Vrms
150 kHz to 80
MHz
6 Vrms
ISM frequency
band
3 V/m
80 MHz to 2.7
GHz
3V
6V
3V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the models
Rooter
®
X3000, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d=1.2×P1/2
d=2×P1/2
d=1.2×P1/2 80 MHz to 800 MHz
d=2.3×P1/2 800 MHz to 2.7 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d Is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey (a.)
should be less than the compliance level in each
frequency range (b.)
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.