Technical specifications
a)
Motor handpiece lithium battery specification
3.7V / 2000mAh
b)
Power adapter specification
Input: ~100V-240V 50Hz/60Hz 0.5-0.2A
Output: DC5V/1A
c)
Motor handpiece mechanical specification
Torque range: 0.4Ncm-4.0Ncm
Speed range: 100rpm-3’300rpm
d)
Wireless charging specification
Frequency range: 112-205KHz
Maximum RF output power of the product: 9.46dBuA/m@3m
1.4
Warnings
Please carefully read this Instruction Manual before first operation.
a)
Do not use this device for anything else than its intended use (refer to Chapter 1.9).
a)
Use original components and accessories only.
b)
Always set torque and speed following the recommendations of the file manufacturer.
c)
Make sure the contra-angle is well connected before to start the motor handpiece (refer
to Chapter 2.1).
d)
Make sure the instrument is well connected and locked before to start the motor
handpiece (refer to Chapter 2.2).
e)
Do not connect or disconnect the contra-angle while the motor is running.
f)
Do not disconnect the instrument while the motor is running.
g)
Be sure to be able to power OFF the device at any time.
h)
Operate and store the device in reliable environment (refer to Chapter 1.7 and Chapter
8).
i)
Do not use the device in the vicinity of fluorescent lamps, radio transmitting devices,
remote control devices, handheld, and mobile high-frequency communication devices.
j)
The motor handpiece, power adapter and charging base are not autoclavable (refer to
Chapter 6).
k)
Replace the lithium battery according to the instructions (refer to Chapter 7.4).
l)
Do not make any changes or modifications to the device. Any change, modification or any
other alteration of the device may violate safety regulations, causing harm to the patient.
m)
In case of motor handpiece frequent overheat, contact a local distributor.
n)
Do not directly or indirectly place the device near heat sources.
o)
Do not cover the device.
p)
Remove the battery of the device in case of long period storage.