3.2.4 Installing the Final Filter
1) Open the attached 0.2 um final filter package.
2) Screw the final filter onto the ultrapure water (UP) outlet till finger tight (no
leaking at dispensing). Do not over tighten it as it may damage threads.
3) Press the UP button to flush out gas in the filter.
4) Press the UP button again to stop dispensing.
If a final ultrafiltration filter for pyrogen removal is needed, please install the
device the same way as the final filter.
3.2.5 Setting Up System Time
System default time is preset. If you need to change the system time, please
follow the procedure below.
1) Press down the START and UP buttons simultaneously to display
system time. Format is Year - Month - Day, and Hour : Minute : Second
2) Press the RO button to enter editing mode.
3) Press the RO button to move the cursor position, the corresponding
number will blink. Press the UP button to increase, and press the RO and
UP buttons simultaneously to decrease values.
4) After editing, press the START button twice to exit the editing mode.
Date & Time
2010-01-01
01 : 01 : 01