www.eu.fishersci.com/fisherbrand
21
2.4.2 Installing the AccuDuo H Pack and U Pack Purification Cartridges
Attention!
These two cartridges work in sequel to ensure high quality of water.
These two packs must be installed in a proper sequence as indicated in the
picture below. Never reverse the positions of these packs.
1) Open the right side panel of the system
2) Remove blue caps on the new cartridges
3) Install H pack to the LEFT side position, and the U pack to the RIGHT
side
4) Wet the O-rings on the cartridge with pure water, gently insert the lower
end first into the opening on the system frame, then down to let the
cartridge sit into its slot
H Pack
U Pack