www.eu.fishersci.com/fisherbrand
18
Prefiltration stand (cartridge based) installation
A. Take out the prefiltration system from the package and put it on a
stable platform or firmly hang it on wall.
B. Connect the venting ball valve and the prefiltration system.
C. Connect tap water and water inlet of
prefiltration support by 3/8 inch
tubing with proper length.
D. Connect water outlet of Prefiltration System and pure water system
by 3/8 inch tubing with proper length.
E. Remove the caps from the cartridges and ports of system. Wet the
O-rings on the cartridge with pure water.
Outlet pressure
gauge
Inlet pressure
gauge
Prefilter cartridge
Water inlet
Cartridge lock-pin
and notch location
Air outlet
Water outlet
Prefiltration
support
Venting ball valve