background image

 

www.eu.fishersci.com/fisherbrand

 

 

20 

 

 

H.  Lock the cartridge with the lock-key to the notch on the look-pin. 

 

 

I.  Pressure  Relief:  Open  venting  ball  valve  connected  with  air  outlet; 

then  turns  the  tap  water  on  gently  to  flush  out  air  in  PreFilter 

Cartridge. Close venting ball valve after degassing. 

 

 

Cartridge 
lock-pin and 
notch location 

Insert the lock-key 
into the notch on 
the lock-pin 

Summary of Contents for 15409699

Page 1: ...www eu fishersci com fisherbrand Accu500 Ultrapure and RO water system USER MANUAL...

Page 2: ...set 15449699 ACCU500 Water System 20L H set 15469699 ACCU500 Water System 30L H set 15419699 ACCU500 Water System 10L H UV set 15439699 ACCU500 Water System 15L H UV set 15459699 ACCU500 Water System...

Page 3: ...Installation 15 3 SYSTEM START UP AND OPERATION 25 3 1 Check List Prior to System Start Up 25 3 2 System Start Up and Time Setting 26 3 3 Routine Operation 32 4 MAINTENANCE 33 4 1 Replacing the Prefi...

Page 4: ...nuous flow of high quality pure water Please contact us or your local distributor if you encounter any issues during installation and use Professional engineers are fully trained to support you Safety...

Page 5: ...RO membrane fouling preventing flush RO membrane and cartridge life detection and many other functions System automatically rinses a new RO membrane User does not need to set special rinse program fo...

Page 6: ...s Pure water can be used in many areas Here are some typical applications With Ultrapure Water With RO Water Important and critical applications Routine and non critical applications HPLC high perform...

Page 7: ...on Rate Ultrapure water Type I 1 5 L min RO water typically 10 to 50 L h at 25 Water Tank Capacity 30 L or 60 L optional RO Rejection Rate 95 Resistivity of Ultrapure Water 18 2 M cm TOC of Ultrapure...

Page 8: ...e of ions then a 0 2 m final filter to the outlet After powering up system goes into operation mode when the START button is pressed If RO water produced does not meet the preset quality requirements...

Page 9: ...compensation range 0 60 to 25 Range of temperature compensation coefficient A channel compensation coefficient setting range 0 5 B channel non linear temperature compensation Display Dot matrix backli...

Page 10: ...90 psi 2 1 3 Tools Needed Not Included Scissors or a box opener to open packages and cut water tubing A wrench to install prefiltration kit 2 2 Items Included Accu500 main system included the followi...

Page 11: ...Flow Diagrams 2 3 1 System Front View 2 3 2 System Control Panel Display Panel Control Panel see explanation on following page RO Outlet UP Outlet Power Indicator RO Indicator UP Indicator START Butt...

Page 12: ...ED after power is turned on RO indicator Above RO Button It turns solid GREEN when dispensing RO water RO indicator light is also used to indicate whether RO water produced meets preset quality parame...

Page 13: ...13 2 3 3 System Back Side View TAP IN FROM TANK TO TANK DRAIN SPARE Socket Fuse Power switch Spare for 2 Pass In from Tap In from Tank Out to Tank RO Drain Socket for the Tank Level Sensor Socket for...

Page 14: ...COMPONENTS A Water inlet B Low pressure switch C RO UP pump D Pressure gauge E Solenoid valve F RO housing G Check valve H Flow restrictor I1 Conductivity sensor I2 Resistivity sensor J Pure water tan...

Page 15: ...ports without unlocking the connector first or use excessive force to pull the stoppers out as this will damage the connector rendering it useless Please follow the instruction below to remove the sto...

Page 16: ...work 1 Install the 3 way ball valve to the tap water inlet Note Fitting is based on 1 2 inch diameter tubing If the tap water outlet is not 1 2 inch you need an adaptor to convert it to 1 2 inch 2 Ins...

Page 17: ...ck Refer to Illustration above Connect the other PE tubing to the water outlet of the Prefiltration pack 5 Put the outlet tubing into a sink Turn on tap water and let it run for a few minutes to clean...

Page 18: ...ter and water inlet of prefiltration support by 3 8 inch tubing with proper length D Connect water outlet of Prefiltration System and pure water system by 3 8 inch tubing with proper length E Remove t...

Page 19: ...9 F Gently insert the lower end first into the opening on the prefiltration support then down to let the cartridge sit into its slot G Push the upper part of the cartridge through the lock pin till ti...

Page 20: ...on the look pin I Pressure Relief Open venting ball valve connected with air outlet then turns the tap water on gently to flush out air in PreFilter Cartridge Close venting ball valve after degassing...

Page 21: ...sequence as indicated in the picture below Never reverse the positions of these packs 1 Open the right side panel of the system 2 Remove blue caps on the new cartridges 3 Install H pack to the LEFT s...

Page 22: ...ight 6 Lock the cartridge with the lock key to the notch on the look pin 7 Install the U Pack cartridge to the RIGHT side of the H Pack in the same procedure as that for H Pack Insert the lock key int...

Page 23: ...h tubing for the tank installation 3 Connect 1 4 inch tubing to the TO TANK on the system and INLET at the base of the PE tank and connect the 3 8 inch tubing to the FROM TANK on the system and OUTLET...

Page 24: ...n the assembled faucet to the tank For the 2 Pass RO UP systems 1 Cut another piece of the1 4 inch 6 35 mm tubing connect the SPARE port on the system and the TO SPARE port on the PE tank base 2 Unwin...

Page 25: ...ults Water quality Municipal water TDS 1000 ppm Water temperature 5 35 Pressure 1 0 6 0 bar 15 90 psi If water hardness is greater than 100ppm Fisher Scientific suggest you to use an external water so...

Page 26: ...llowing order Year Month Day then Hours Minutes Second row RO status XX X S cm 25 RO Water conductivity compensated to 25 Third row UP status XX X M cm 25 Ultrapure UP water resistivity compensated to...

Page 27: ...dispensed through the system RO DISPENSE RO water dispense mode time displayed is the preset time of dispensing if set countdown UP DISPENSE UP water dispense mode time displayed is the preset time of...

Page 28: ...utomatically discharges failed RO water to the drain instead of the water tank The RO indicator light blinks It is normal that RO water is above the upper conductivity limit at system start up It usua...

Page 29: ...m goes to standby 3 2 2 2 Start Up the System When the START button is pressed system enters operational mode after an automatic flush of the RO membrane for 180 seconds The System automatically detec...

Page 30: ...an excessive amount of water is dispensed water level inside the PE tank may fall below the minimum set point Water dispensing is stopped automatically System will display Tank Empty on the display p...

Page 31: ...se install the device the same way as the final filter 3 2 5 Setting Up System Time System default time is preset If you need to change the system time please follow the procedure below 1 Press down t...

Page 32: ...panel displays Tank Full System will automatically restart once water is being dispensed 3 Dispensing Water 1 RO Water Press the RO button Press once to dispense RO water press again to shut RO valve...

Page 33: ...er minute H Pack U Pack Protects the system 12 months Reduce in resistivity of ultrapure water Polypropylene PP Prefilter Remove particles 2 4 weeks Reduced feed water flow Softening Cartridge Remove...

Page 34: ...ted filters 5 Install new filters the 10 m filter in the upstream housing the activated carbon AC cartridge in the middle housing the 1 m filter in the downstream housing 6 Mount the filter housing ba...

Page 35: ...to put system into standby Power off the system 2 Remove the system right side panel 3 Remove the used cartridges by lifting up the lock chip then pull the cartridges out 4 Remove the blue caps on the...

Page 36: ...o dispense UP water for 3 5 minutes until the UP indicator light stops blinking and no bubbles come out of the spout 8 Press UP button again to stop dispensing 9 Screw a new final filter to the UP ter...

Page 37: ...p cover to prevent UV light leakage This cover must be on at ALL TIMEs when a UV lamp is installed Caution Keep the UV lamp straight in and out of the stainless steel chamber during its installation t...

Page 38: ...n contact with the quartz glass of the UV lamp 7 Carefully insert the new UV lamp into its chamber When about 2 3 of the lamp is inserted hold the UV lamp and connect it to the ballast cable connector...

Page 39: ...nt of water If you still feel uncomfortable please seek medical attention Caution RO reverse osmosis membrane should be replaced by a fully trained professional technician 1 Press the START button to...

Page 40: ...ts at joints with water tubing and then pull tubing out in a gentle manner 6 Connect labeled tubing inside the system to the matching labels RO IN RO OUT and RO DRAIN on the new membrane housing 7 Ins...

Page 41: ...dicated in the picture below 1 Unplug the power cord from the system 2 Put the head of a small head screw driver in the notch 3 Use the back wall of the power cord receptacle as the point of support t...

Page 42: ...bing connected to the From Tank port put it into the sink Note Please check local regulations on how to properly discharge 0 1 M NaOH 6 Open the blue valve underneath the tank drain all solution 7 Clo...

Page 43: ...ace AccuDuo H and U Pack cartridges Low Pressure Alarm No inlet water Check tap water faucet Prefilter clogged Replace prefilters in the Prefiltration pack Pressure at the tap water too low Call servi...

Page 44: ...d Replace pre filters Reverse osmosis RO membrane clogged Replace RO membrane Final filter clogged Replace final filter Water tank empty Wait till the water tank is filled UP Resistivity dropped at di...

Page 45: ...stem 15L H set 15449699 ACCU500 Water System 20L H set 15469699 ACCU500 Water System 30L H set 15419699 ACCU500 Water System 10L H UV set 15439699 ACCU500 Water System 15L H UV set 15459699 ACCU500 Wa...

Page 46: ...n 12 pk Pack 15489639 Activated carbon cartridge Activated carbon cartridge 10 inches 12 pk Pack 15499639 Softener cartridge Cation softener cartridge 10 inches 12 pk Pack 15469669 Reverse osmosis mem...

Page 47: ...er Low TOC Set 15469689 UV lamp 185 254 nm dual wavelength UV lamp Each 15409659 Final filter 0 2 m final capsule filter 1 pk Each 15409689 Tank vent filter Tank Vent Filter with CO2 Remover Each 1541...

Page 48: ...110V 60 Hz RO Booster Pump for Accu500 10 RO Booster Pump for Accu500 20 RO Booster Pump for Accu500 30 UP circulation pump Resistivity Sensor Conductivity Sensor Low Pressure switch Pressure Gauge Co...

Page 49: ...erials and workmanship when used in compliance with instructions and operating conditions specified in this manual Thermo Fisher Scientific warrants the system for 24 months from the earlier of 1 The...

Reviews: