![Fisher Scientific Isotemp 800 Series Manual Download Page 24](http://html1.mh-extra.com/html/fisher-scientific/isotemp-800-series/isotemp-800-series_manual_2296654024.webp)
The following sections describe procedures for servicing
the 800 Series ovens. The first procedure, Replacing the
Door Gasket
, and the last,
Replacing the Handle and
Hinges
, may be performed by most users. However, all
other service procedures involve possible exposure to line
voltage. Only qualified service personnel should under-
take these procedures. The second section,
Accessing
the Electronics Compartment
, describes procedures
required for subsequent service sections and is refer-
enced by these later sections when required.
In the event service is required beyond that available by
the customer, or for warranty service, contact Fisher
Service Dept. at 1-800-395-5442.
For technical assistance call 1-800-926-0505.
Replacing the Door Gasket
The Isotemp 800 Series ovens incorporate a durable,
fiberglass door gasket to minimize heat loss. Should the
gasket become defective or be damaged, it may be
replaced by following the procedures below.
1.
Set the power switch to the off position and
open chamber door.
2.
Remove the screws securing the stainless steel
inside door panel.
3.
The gasket is attached to the inner door panel
by screws passing through the gasket lip and
the sandwich of the inner panel and door.
4.
Remove the old gasket and position the new
gasket lip under the stainless steel inner panel.
Push the screws through the gasket and replace
the inner panel. Trim gasket length as neces-
sary.
24
Service
Warning
Service procedures requiring access
to the electronics compartment involve
exposure to line voltage and should be
done only by qualified service person-
nel.
Disconnect oven from power
source before attempting repairs.
Warning
Allow oven to cool to ambient tem-
perature before attempting repair.
Summary of Contents for Isotemp 800 Series
Page 34: ...34 Schematic Diagram...
Page 35: ...35...