Actual Display to indicate the desired step num-
ber to jump to. Press the Menu Key to enter the
jump step. NOTE: The jump step number must
be lower than the current step number. The
jump cannot be to the current step or a higher
step.
19. The Set Display then shows “JC” to indicate that
the operator must select the number of times
the jump loop is to be repeated. Press the
Menu Key to enter the jump count.
20. The displays indicate “no rtn”.
If further steps
are to be programmed
, press the Menu Key to
continue at the “no return” prompt. This will
advance program editing to the next step num-
ber. Continue setting up the program at step 5
of this procedure.
21.
If no further program steps are required,
press
the Up or Down Arrow Key. The “yes return”
prompt will appear. Press the Menu Key to
return to Control mode operation.
22.
When program editing is complete
the controller
will then prompt to save the program, “1 Save”.
Press the Menu Key to save the program, or
use the Up or Down Arrow Key to change the
Actual Display to NO, which will leave the cur-
rent program as it was found.
To initiate the program, see
Running a Program
.
Running, Stopping or Resuming
a Program
Prior to running the program in memory, first make certain
that the current set temperature is the temperature from
which the program is to start. For example, a program
intended to ramp the oven temperature from ambient to
250°C, if started with the current set temperature at
300°C, will actually ramp the temperature down from
300°C to 250°C.
14
O
PERATION