5.
Press the Menu Key to edit step #1, or use the
Up or Down Arrow Key to change the step num-
ber to the desired step for editing before press-
ing the Menu Key.
6.
The Set Display then indicates “StyP” to signify
that step type must be selected from a menu of
four operations being shown in Actual Display.
The four menu selections are Set Point, Soak,
End or Jump Loop as shown to the left. Use the
Up/Down Arrow Keys to change the Actual
Display to the desired step type. Press the
Menu Key to enter the desired step type. NOTE:
For step 1 in a program, the jump loop option is
not available.
7.
If the desired program step is an End
, press
the Menu Key to select it when End is in the
Actual Display. Otherwise, skip to step 10.
8.
The display shows one of the two options for an
End step. To view the selections, press either
the Up or Down Arrow Key. The End command
signifies the final step in a program.
9.
If the final temperature is to be held indefinitely
at the end of the ramp and soak program, press
the Menu Key when “HoLd” is displayed on the
Actual Display. If heater power is to be turned
off at the end of the program, press the Menu
Key when Actual Display indicates “OFF”. Skip
to step 22.
10.
If the desired program step is a ramp to a Set
Point
, press the Menu Key to select it when the
Actual Display reads “SP.” Otherwise, skip to
step 13.
11. If a ramp to a Set Point is selected, the Set
Display first indicates “SP” to signify that the
operator must select a final target temperature,
or set point. Use the Up or Down Arrow Key to
adjust the Actual Display to indicate the desired
final temperature for the ramp. Press the Menu
Key to enter the selected set point temperature.
12
O
PERATION