12. The Set Display then indicates “rAtE” to prompt
the operator to select a desired heating or cool-
ing rate in degrees/min. Press and hold or until
the desired rate is shown in the Actual Display.
Press the Menu Key to enter the selected rate.
Proceed to step 20.
13.
If the desired program step is a Soak,
press the
Menu Key to select it when the Actual Display
shows “SoAH.” Otherwise, skip to step 17.
14. If a Soak is selected, the Set Display first indi-
cates “HOUr” to signify that the operator must
select the number of hours that the current set
temperature will be maintained. Use the Up or
Down Arrow Key to adjust the Actual Display to
indicate the desired number of soak hours (max-
imum of 23). Press the Menu Key to enter the
selected number of soak hours.
15. The Set Display then shows “Min” to indicate
that the operator must select the number of
additional minutes for which the current set tem-
perature will be held. Use the Up or Down
Arrow Key to set the Actual Display to the
desired number of soak minutes (maximum of
59). Press the Menu Key to enter the selected
number of soak minutes.
16. The Set Display then shows “SEC” to indicate
that the operator must select the number of
additional seconds for which the current set
temperature will he held. Use the Up or Down
Arrow Key to adjust the Actual Display to indi-
cate the desired number of soak seconds (maxi-
mum of 59). Press the Menu Key to enter the
selected number of soak seconds. Proceed to
step 20.
17.
If the desired program step is a Jump Loop
,
press the Menu Key to select it when the Actual
Display shows “JL.”
18. If a Jump Loop is selected, the Set Display first
indicates “JS” to signify that the operator must
select the step number destination for the jump.
Use the Up or Down Arrow Key to adjust the
13