behr TRS 300
– 28 –
271005
behr TRS 300
– 29 –
© behr Labor-Technik 2007
271005
Code Meaning
Remedy
010
The digester‘s power-supply
cable is not connected to the
TRS 300 or the plug is not firmly
inserted
Switch the TRS 300 off. Insert the power-supply cable
connector of the digester firmly into the „Heating“ socket
(Page 7, 2.4) on the rear panel of the TRS 300 and secure
it using the union nut. Then switch the unit on again.
The digester‘s thermal cut-out has
tripped
Switch the unit off and allow the digester to cool down
for a while. Check that the Function % has not been
inadvertently selected in your temperature program.
The digester‘s heating system
is defective or the triac control
system of the TRS 300 is
defective
The digester and/or the TRS 300 must be repaired
or replaced. Switch off both units and inform the behr
after-sales service.
011
The unit heats without stopping
The TRS 300 is defective. Switch it off and inform the behr
after-sales service.
100
General error
Switch the unit off and inform the behr after-sales service.
If the TRS 300 cannot be switched on (display remains dark), the mains fuse may have tripped. Switch
the unit off and on again after a few minutes. If the unit can still not be switched on or if the problem oc-
curs repeatedly, inform the behr after-sales service.