C/M function: Heating rate
(degrees Celsius per Minute)
In this function, the unit heats up uniformly to the
specified
Temperature
(in degrees Celsius) in the
specified
Time
(in minutes). It thus operates on a
temperature gradient.
The temperature gradients which can in fact be
operated depend on the type and heat output
of the digester, and also on the heat capacity
of the samples.
Perform a trial run and find out for yourself,
using the temperature display, whether the
temperature plots set can actually be achieved.
Function E: End
This function is used to inform the TRS 300 that
the temperature program ends here.
Every program memory slot consists of ten slots
for program steps. The TRS 300 works through
these ten program steps in sequence if it does not
previously encounter an End step.
The last program step of every program must
therefore be an End step.
An End step is only
not needed if the program actually has the maxi-
mum of ten steps.
TIME
FUNCTION
TEMPERATURE
Program
01/01
<<==
==>>
350
120
C/M