2-7
Trimming the analog output is a calibration of the output circuitry, by setting
the 4 and 20 mA points. Once the 4 and 20 mA points are set, all
intermediate values are automatically adjusted. See Menu Trees in Section
3 for a corresponding menu tree.
From the Online menu, select Device setup.
From the Device setup menu, select Diag/
Service.
From the Diag/Service menu, select
Calibration.
From the Calibration menu, select Trim
analog output.
From the Trim Analog Output menu, select
the desired option and follow the on-line
instructions.