C-3
NOTE: I
m
= I
scBarrier
+ I
sc275
= I
scBarrier
+ 32
m
A
(
m
A)
GROUPS
A & B
GROUP
C
GROUP
D
170
0.80
5.50
9.90
160
1.00
6.20
11.20
150
1.30
7.00
12.70
140
1.60
8.00
14.60
130
2.00
9.00
16.90
120
2.50
10.00
19.80
110
3.00
12.00
23.60
100
4.00
15.00
28.50
90
5.00
18.00
35.10
85
5.50
20.00
39.30
80
6.00
22.00
44.40
75
6.70
25.00
50.50
70
7.50
28.00
57.90
65
8.80
34.00
67.10
62
9.50
37.00
73.70
60
10.00
40.00
78.70
57
11.00
43.00
87.10
55
12.00
48.00
93.50
50
15.00
56.00
113.10
45
19.00
70.00
139.40
40
23.00
87.00
176.30
35
31.00
110.00
229.90
32
36.00
135.00
274.80
30
40.00
150.00
312.40
28
46.00
170.00
358.40
25
58.00
210.00
449.00
23
68.00
250.00
530.10
21
82.00
300.00
635.30
20
90.00
330.00
700.00
Table 2: Maximum Allowable Connected Inductance
I
m
La (
m
H)
00275-0081
DWG No.
Rev. D.
Sheet 1 of 2