24
Dimensions are nominal and illustrations and specifications are based
on the latest product information available at the time of publication.
The right is reserved to make changes at any time without notice.
SAFEAIRE
®
LABORATORY FUME HOODS
OPERATION, MAINTENANCE AND INSTALLATION INSTRUCTIONS
MAINTENANCE & ADJUSTMENTS
Stainless Steel
Edge Protector
SASH GLASS REPLACEMENT – RESTRICTED BYPASS AND CONSTANT VOLUME BENCH TOP FUME HOODS
THAT HAVE BEEN INSTALLED
1. Remove louver panel by removing one screw in lower
corner at each end.
2. Pull bottom out and lift entire panel up to remove.
3. Remove both front corner posts, See Page 22.
WARNING
If cable is frayed or damaged, it MUST be replaced to
avoid personal injury or damage to the fume hood
4. After removal of one of the corner posts, clamp the sash
cable to the top of the side frame assembly. See page 23.
5. Remove screws that hold the sash guide to the front frame
upright and slide the guide up and away from the sash.
Care should be taken when the sash guide is removed so
that the glass does not fall from the remaining guide.
Remove
two screws
2
1
Right Sash Guide
Sash Glass
Louver Panel
Right
Corner Post
Left
Corner Post
Bottom Sash Frame
Left Sash Guide
Fume Hood
Frame
Assembly
Summary of Contents for SAFEAIRE series