
23
OSR-CWDM Users Guide Revision A 1/2019
Section 11
Video Scope Operation
Centering a Connector Image
Once the image is stable and focused, use the stylus to touch the approximate center of the connector to snap
the image near the center of the display and the center of the grading rings. Use the LRUD buttons to fine tune
the image to the center of the display.
Pausing Image Scan
To freeze an image in position and focus level for inspection, press the scan button. This will pause the image at
the current position and focus level. Simply press the scan button again to set the operational mode back to live
mode for focus and position adjustments
Grading Rings
Turn on or off the grading rings by pressing menu and using the left or right buttons to highlight the grading ring
icon and press select or touch the icon on the touch screen to toggle the rings on and off.
Pass/Fail
Using the pass/fail criteria tables on the next page, determine if the connector passes the IEC61300-3-35
standard. There is a guide to contamination size located at the bottom of the image. To mark a connector as
Pass or Fail, enter the menu mode, highlight the Pass/Fail Label icon and press select to cycle through pass, fail,
or no grading. The touch screen may also be used to cycle through these settings. Once the pass fail status has
be marked, the scan may be paused with the scan button and the points of contamination that caused a failure
may be marked on the image.
Marking Points of Contamination
To mark the points of contamination the unit must be in paused mode. Pick up the contamination marker by using
the stylus and touching the contamination sample size that is required. Touch the image to place the marker.
The marker may be fine tuned with the stylus to cover the contamination point on the image to be marked. Once
the marker is positioned properly, touch the area just below the image that states, "Place marker, touch here to
apply" to lock the marker in place. Repeat this as necessary to mark all the points that need to be indicated for
the pass/fail status. At this point the image should be save with the quick save icon to record the information.
NOTE:
Marking the points of contamination should be done last, just before saving the image. Once the unit is
returned to live mode the markers are removed.
Exiting Video Scope Operation
Enter the menu mode, use the left or right buttons to highlight the Home icon and press select or touch the home
icon directly on the touch screen.
Note
The majority of function and features in the Video Scope are operable with the touch screen. It
is not necessary to enter the menu mode when using the touch screen. The menu mode needs
to be entered when the focus of the LRUD buttons needs to be taken off the positioning feature
and moved to the icon cursor.