Section 11
Video Scope Operation
Quick Save
Press the menu button, use the left or right buttons to highlight the quick save icon and press the select button or
touch the quick save icon to store a file image of the scope screen to the next location.
Quick Load
Press the menu button, use the left or right buttons to highlight the quick load icon and press the select or touch
the quick load icon on the touch screen to load the next stored scope image.
File Management
To open file management, press the menu button, use the left or right buttons to highlight the file management
(Folder) icon and press the select button or touch the file manager icon on the touch screen. This will open the
standard file management screen as described in Section 10 of this User’s guide. Please refer to this section for
information on file processing.
Help
Press the menu button, use the left and right buttons to highlight the help icon and press select or touch the help
icon on the touch screen to view the help information.
11.3 Video Scope Operation Menu
The following functions are only available for use when the video scope is in live scan mode. There are two
modes in the video scope, menu mode and live mode. The menu button toggles between the two modes.
Touching one of the icons will also open the menu mode. When in menu mode there will be a highlight around
one of the icons on the top or the bottom of the display.
Note
Touching icons on the video scope screen will put the unit into menu mode and activate or
toggle the function. To exit menu mode it is necessary to use the menu button on the keypad.
Grading Rings Icon
Used to turn on and off the rings that indicate the IEC61300-3-35 grading zones.
By entering the menu mode with the menu button, use the LRUD buttons to highlight the pass/fail grading rings
overlay icon and press the select button or touch the icon. Exit the menu mode and use the LRUD buttons to
center the image in the rings. Use the appropriate table from Section 11.5 to grade the connector end face. 2, 3,
5 and 10 micron contaminant examples are displayed just below the connect image on the video scope display.
Pass/Fail Label Icon
Used to cycle through pass, fail or no grading (off) of the pass/fail/auto indicator. To set the pass/fail indicator,
press the menu button, use the left or right buttons to highlight the pass/fail icon and press select to cycle through
the pass/fail states, or touch the icon.