background image

PAGE / 14  

Doc. P/N 06-905 

Rev. 1 / December 2018 

 

7. 

The current reading displays, along with “ARM PASSES” or “FAIL REPLACE ARM.” 

 

 

 

NOTE: The upper limit of each test range is provided to help you determine when a component is nearing the 
end of its useful life and should be replaced. For the Fire ARM, the average of the four test readings must be 
greater than or equal to 95% of 3.5 amps, or 3.325A. If the average is approaching 3.325A, consider replacing 
the ARM. 

8. 

Press and release the ENTER button to repeat the test, or press and release the SELECT and ENTER buttons 
at the same time to return to the Main menu. 

9. 

Power down and remove the tester.   

10. 

End of testing. Place system back into normal operational state.     

10.0

 

EP ARM TESTING 

10.1.

 

EP ARM Operational Overview 

The Output Analyzer can be used to perform two important testing functions associated with the EP ARM’s firing 
circuit.  The  first  testing  function  assesses  the  EP  ARM’s  ability  to  supply  the  correct  current  (amps)  required  to 
activate the actuator devices connected to the firing circuit; the second testing function assesses the integrity of the 
field wiring by measuring the supervisory current at the EP-ARM. 

When the device is connected to an EP ARM for testing, the activation signal is provided from a pressure or optical 
detector. The device will then measure the current supplied by the EP-ARM’s firing circuit and display the result on 
the LCD panel. Three readings will be taken to ensure the EP ARM is supplying adequate firing current, with readings 
occurring instantaneously. 

10.2.

 

Interpreting EP ARM Measured Results 

Once you’ve used the Output Analyzer to test the activation current supplied by the EP-ARM, the Analyzer will display 
the results on the front panel as either EP ARM PASSES or FAIL REPLACE ARM.   

When the Analyzer is used to test the supervisory current supplied by the  Explosion Protection Controller, it will 
display the supervisory current reading in milliamps. The acceptable supervisory current reading at the  EP-ARM is 
20 – 54mA.  

 

 

1.78 2.2 2.6 2.8 
FAIL REPLACE ARM 

5.65 6.1 3.9 2.3 
ARM PASSES 

Summary of Contents for 10-2983

Page 1: ...SOLUTIONS Fire Protection Explosion Protection Overpressure Protection Pressure Activation Doc P N 06 905 Rev 1 December 2018 PRODUCT MANUAL Output Analyzer P N 10 2983...

Page 2: ......

Page 3: ...R FITNESS FOR A SPECIFIC PURPOSE Refer to www fike com terms conditions for Fike s full TERMS AND CONDITIONS OF SALE TERMS OF USE Do not alter modify copy or otherwise misappropriate any Fike product...

Page 4: ......

Page 5: ...urrent Test 8 7 3 3 Test 3 EPC Field Test 9 7 3 4 Test 4 Field Wiring Supervisory Current Test 10 8 0 IRM TESTING 11 8 1 IRM Operational Overview 11 8 2 Interpreting IRM Measured Results 11 8 3 IRM Te...

Page 6: ...ng Circuit Activation Current 6 Figure 7 3 2 1 EPC Series Firing Circuit Supervisory Current Test 8 Figure 7 3 3 1 Field Wiring Activation Current Test 9 Figure 8 3 1 IRM Circuit Activation Current Te...

Page 7: ...er EPC P N E10 0066 Impulse Release Module IRM P N 10 2748 Fire Agent Release Module Fire ARM P N 10 1832 Explosion Protection Agent Release Module EP ARM P N E10 010 The device s portability and ease...

Page 8: ...ing cord but does not come with a wall charger or a 1 Ohm Resistor used in EPC testing Please obtain a charger to fully charge the battery before testing To ensure this device remains compliant with E...

Page 9: ...r modifications not expressly approved by the party responsible for compliance could void the user s authority to operate this equipment This product complies with the essential requirements of the fo...

Page 10: ...attery charge is displayed on initial power up and you can check it periodically by using Option 7 Chk Batt Chrg to keep track of usage 6 0 CONTROLS INTERFACE On initial power up the battery charge wi...

Page 11: ...ttery charge is displayed on initial power and can be viewed at any time under Option 7 of the main menu 7 0 EPC TESTING 7 1 EPC Operational Overview The Output Analyzer is used to perform two importa...

Page 12: ...he end of its useful life and should be replaced For the EPC the average of the four test readings must be greater than or equal to 95 of 3 amps or 2 85A If the average is approaching 2 85A consider r...

Page 13: ...gs obtained during the test EPC PASSES REPLACE EPC will display on the front panel of the Analyzer See Section 7 2 Interpreting EPC Measured Results for more information or if you have questions regar...

Page 14: ...lyzer wait the three seconds for the battery charge status to flash then press and release the SELECT button to arrive at the EPC Tool menu Option 1 press the ENTER button to choose this option The su...

Page 15: ...are disconnected except at the field box under testing 4 At the field box under testing install short wire leads into the GCA firing circuit terminals 5 On the first field box to be tested use the 14...

Page 16: ...zer test verifies the supervisory current being supplied by the EPC to the field devices through series firing circuit field wiring The Analyzer comes with jumpers and test leads used to perform this...

Page 17: ...upplying adequate firing current with readings occurring momentarily there is a slight delay between fire control panel activation and firing circuit release IMPORTANT The Output Analyzer ACTIVATE but...

Page 18: ...me 4 Wait 60 seconds to allow the capacitors time to charge If the panel does not immediately return to normal reset the panel to a normal state 5 Remove the 0 Shunt from the Analyzer The supervisory...

Page 19: ...about 10 min then shunt all GCA s per 06 106 ARM III Rev 2 connected to the system before proceeding 2 Connect the Analyzer to the ARM as shown in Figure 9 3 1 below Figure 9 3 1 Fire ARM Circuit Act...

Page 20: ...upply the correct current amps required to activate the actuator devices connected to the firing circuit the second testing function assesses the integrity of the field wiring by measuring the supervi...

Page 21: ...Ready for Test Reset the main EP Control Panel if required to obtain a normal panel and wait 60 seconds for the capacitor to charge 6 Activate the EP ARM by initiating a detection input that will caus...

Page 22: ...with any battery operated device the battery in the Output Analyzer should be charged periodically using a typical USB C 5Vdc connection Ensure the voltage rating of the connection is 5Vdc maximum or...

Reviews: