background image

PAGE / 12  

Doc. P/N 06-905 

Rev. 1 / December 2018 

 

9.0

 

   FIRE ARM TESTING 

9.1.

 

Fire ARM Operational Overview 

The Output Analyzer can be used to perform two important testing functions associated with the Fire ARM’s firing 
circuit.  The first  testing function assesses the Fire ARM’s ability to supply the correct current  (amps) required to 
activate the actuator devices connected to the firing circuit; the second testing function assesses the integrity of the 
field wiring by measuring the supervisory current at the ARM. 

When the Analyzer is connected to a Fire ARM for testing, the activation signal is provided via a release event on the 
fire alarm control panel, such  as when a release station is activated. The  Analyzer will then measure the current 
supplied by the ARM firing circuit and display the result on the LCD panel. Four readings will be taken to ensure the 
Fire ARM is supplying adequate firing current, with readings occurring momentarily (there is a slight delay between 
fire control panel activation and firing circuit release).  

IMPORTANT: The ACTIVATE button does not apply to Fire ARM testing. Instead, activate using a release event on the 
fire alarm control panel (e.g., activate release station). 

3. 

Enable the release circuit and IRM with the Analyzer installed (one per IRM or repeat testing for each one at 
a time).  

4. 

Wait 60 seconds to allow the capacitors time to charge. If the panel does not immediately return to normal, 
reset the panel to a normal state.  

5. 

Remove the 0Ω Shunt from the Analyzer. The supervisory current will be displayed (.20mA - .54mA). 

 

 

6. 

Record the IRM supervisory current reading for future reference, since subsequent tests will overwrite current 
test data. 

7. 

Install the 0Ω Shunt (short jumper) onto the Analyzer 

8. 

The  display  will  read  “Ready  for  Test.”  Reset  the  panel  if  required  to  obtain  a  normal  panel,  and  wait  60 
seconds for the capacitor to charge. 

 

 

9. 

Activate the IRM by initiating an input that will cause the panel output to trigger the IRM. 

10.  The current reading will be displayed, along with “IRM PASSES” or “FAIL REPLACE IRM.” 

 

 

 

NOTE: The upper limit of each test range is provided to help you determine when a component is nearing the 
end of its useful life and should be replaced. For the IRM, the average of the four test readings must be greater 
than or equal to 95% of 4.5 amps, or 4.275A. If the average is approaching 4.275A, consider replacing the IRM. 

11.  Record the IRM firing current reading for future reference, since subsequent tests will overwrite current test 

data. 

12.  Press and release the ENTER button to repeat the test or press and release the SELECT and ENTER buttons at 

the same time to return to the Main Menu. 

13.  End of testing. Place system back into normal operational state.     

IRM Supervisory 
Current .32mA 

IRM ANALYZE 
Ready for Test 

2.39 3.4 4.0 3.5 
FAIL REPLACE IRM 

7.34 6.7 4.7 3.1 
IRM PASSES 

Summary of Contents for 10-2983

Page 1: ...SOLUTIONS Fire Protection Explosion Protection Overpressure Protection Pressure Activation Doc P N 06 905 Rev 1 December 2018 PRODUCT MANUAL Output Analyzer P N 10 2983...

Page 2: ......

Page 3: ...R FITNESS FOR A SPECIFIC PURPOSE Refer to www fike com terms conditions for Fike s full TERMS AND CONDITIONS OF SALE TERMS OF USE Do not alter modify copy or otherwise misappropriate any Fike product...

Page 4: ......

Page 5: ...urrent Test 8 7 3 3 Test 3 EPC Field Test 9 7 3 4 Test 4 Field Wiring Supervisory Current Test 10 8 0 IRM TESTING 11 8 1 IRM Operational Overview 11 8 2 Interpreting IRM Measured Results 11 8 3 IRM Te...

Page 6: ...ng Circuit Activation Current 6 Figure 7 3 2 1 EPC Series Firing Circuit Supervisory Current Test 8 Figure 7 3 3 1 Field Wiring Activation Current Test 9 Figure 8 3 1 IRM Circuit Activation Current Te...

Page 7: ...er EPC P N E10 0066 Impulse Release Module IRM P N 10 2748 Fire Agent Release Module Fire ARM P N 10 1832 Explosion Protection Agent Release Module EP ARM P N E10 010 The device s portability and ease...

Page 8: ...ing cord but does not come with a wall charger or a 1 Ohm Resistor used in EPC testing Please obtain a charger to fully charge the battery before testing To ensure this device remains compliant with E...

Page 9: ...r modifications not expressly approved by the party responsible for compliance could void the user s authority to operate this equipment This product complies with the essential requirements of the fo...

Page 10: ...attery charge is displayed on initial power up and you can check it periodically by using Option 7 Chk Batt Chrg to keep track of usage 6 0 CONTROLS INTERFACE On initial power up the battery charge wi...

Page 11: ...ttery charge is displayed on initial power and can be viewed at any time under Option 7 of the main menu 7 0 EPC TESTING 7 1 EPC Operational Overview The Output Analyzer is used to perform two importa...

Page 12: ...he end of its useful life and should be replaced For the EPC the average of the four test readings must be greater than or equal to 95 of 3 amps or 2 85A If the average is approaching 2 85A consider r...

Page 13: ...gs obtained during the test EPC PASSES REPLACE EPC will display on the front panel of the Analyzer See Section 7 2 Interpreting EPC Measured Results for more information or if you have questions regar...

Page 14: ...lyzer wait the three seconds for the battery charge status to flash then press and release the SELECT button to arrive at the EPC Tool menu Option 1 press the ENTER button to choose this option The su...

Page 15: ...are disconnected except at the field box under testing 4 At the field box under testing install short wire leads into the GCA firing circuit terminals 5 On the first field box to be tested use the 14...

Page 16: ...zer test verifies the supervisory current being supplied by the EPC to the field devices through series firing circuit field wiring The Analyzer comes with jumpers and test leads used to perform this...

Page 17: ...upplying adequate firing current with readings occurring momentarily there is a slight delay between fire control panel activation and firing circuit release IMPORTANT The Output Analyzer ACTIVATE but...

Page 18: ...me 4 Wait 60 seconds to allow the capacitors time to charge If the panel does not immediately return to normal reset the panel to a normal state 5 Remove the 0 Shunt from the Analyzer The supervisory...

Page 19: ...about 10 min then shunt all GCA s per 06 106 ARM III Rev 2 connected to the system before proceeding 2 Connect the Analyzer to the ARM as shown in Figure 9 3 1 below Figure 9 3 1 Fire ARM Circuit Act...

Page 20: ...upply the correct current amps required to activate the actuator devices connected to the firing circuit the second testing function assesses the integrity of the field wiring by measuring the supervi...

Page 21: ...Ready for Test Reset the main EP Control Panel if required to obtain a normal panel and wait 60 seconds for the capacitor to charge 6 Activate the EP ARM by initiating a detection input that will caus...

Page 22: ...with any battery operated device the battery in the Output Analyzer should be charged periodically using a typical USB C 5Vdc connection Ensure the voltage rating of the connection is 5Vdc maximum or...

Reviews: