F12
Auto Tare Function
** Off
On
Contin
Notes: -
•
Off
= Auto Tare Function disable.
•
On
= Only the first table weight applied will be tared off. Minimum tare load ≥ 2d.
•
Contin
= All stable weight applied will be tared off. Minimum tare load ≥ 20d.
•
If Contin is set, select also delay time (0.0 ~ 9.9 second). Delay time is the time duration
from a stable weight is detected and until it is automatically tare off.
•
Delay time value is displayed on LCD3. Enter the preferred delay time value through
numeric keys and then press
[Print/M+]
to save.
F13
Repetitive Tare
Function
Off
** On
If F12 is set = Contin, Repetitive Tare setting “Off” will be surpassed.
F14
Operation Buzzer
Kb (keypad buzzer)
(
**On
/Off)
St (System buzzer)
(
**On
/Off)
F15
Check Buzzer &
Action on Negative
Value
oFF
** in
out
Hi
Lo
oFF
= Check Buzzer disabled.
in
= Check Buzzer activates when reading is within range.
out
= Check Buzzer activates when reading is out of range.
Hi
= Check Buzzer activates when reading more than Hi limit.
Lo
= Check Buzzer activates when reading lower than Lo limit.
Action on Negative Value
Off = Check Mode Disable when value is less than zero.
** On
= Absolute value, all negative values will be deemed and checked as positive
ones.
F16
UART (Comport)
Setting
4 UARTs (UART 0, 1, 2, 3) are available. Refer to
paragraph 8
for more
setting details.
F18
Machine ID and
Group Number
Machine ID
Group Number
Notes: -
a.
Machine ID number entered is also used as the address number of the instrument for
RS485 communication.
b.
When machine ID is = blank. It means that this instrument does not have any address
number in the RS485 network.
c.
In multi-instrument application with the same RS485 network, each machine should have
a unique machine ID number.
d.
Group number can be considered as production line number in multi-instrument
application.
e.
The machine ID and group number entered will be carried out automatically on output
formats Lab 2 ~ Lab 5.
Setting procedures: -
1.
Go to F18 and then press
[Print/M+]
.