Go to fetco.com for the latest versions of all information
Page 7
User Guide & Operator instructions
P211
e0
November 2021
B General
Brewer Operation Control Settings, Adjust Brew Flow Rate
POSITION
Program Items
Factory set Default
Programming Range Increments
Notes
B1
Tank Temp.
200°F-or-93° C
NOTE: Equipment is
Fahrenheit by default
77° to 97°C
170°F to 207°F
0.5°C
1.0°F
Chart to correct for
high altitude below
B2
Brew at Temp.
“YES”
ON/OFF
YES/NO
SEE NOTE BELOW
B4
Show Tank
Temperature
YES
YES/NO
To display HW tank
temperature on screen
B5
Units of Measure
TEMPERATURE
° Fahrenheit
Fahrenheit/Celsius
C/F
NOTE
: Overwrites
user settings
(see page 9)
B6
Units of Measure
VOLUME
Liters
Gallons/Liters/Ounces
L/Gal/Oz
NOTE
: Overwrites
user settings
(see page 9)
B7
Customer Name
Off
NO or YES
For name on screen
B7
Customer Name
(only if above is “ON) Scroll with batch keys A-Z;a-z;0-9
16 characters total
B9
Demo Mode
DEMO ON/OFF
Demonstrates the controls
for training. Disables all
components in demo mode
B10
Eco Mode
On
ON/OFF
YES/NO
If Selected: Lowers hot
water tank temperature after
preset time of inactivity
B11
Eco Idle Time
1Hr
1-6 hours
1 hour
Time of inactivity to go into
ECO Mode
B12
Eco Idle Temp
170°F
158-176°F
1 degree
Temperature that hot water
tank is lowered to when in
ECO Mode
B13
Filter Life
OFF
ON/OFF
YES/NO
Water filter life is accessed
in G-Counters. This is user
set and will display indicator
to change water filter
Chart to correct for altitude for boiling point
in tank water temperature.
[ft]
[m] Suggested
Setting[°F]
Boiling
point[°F]
Suggested
Setting[°C]
Boiling
point [°C]
0
0
205
212.0
96
100.0
500 152
205
211.0
96
99.5
1000 305
200
210.1
93
98.9
2000 610
200
208.1
93
97.8
2500 762
200
207.2
93
97.3
3000 914
200
206.2
93
96.8
3500 1067
197
205.3
92
96.3
4000 1219
195
204.3
91
95.7
4500 1372
194
203.4
90
95.2
5000 1524
194
202.4
90
94.7
5500 1676
193
201.5
89
94.2
6000 1829
192
200.6
89
93.6
6500 1981
191
199.6
88
93.1
7000 2134
190
198.7
87
92.6
7500 2286
188
197.8
86
92.1
8000 2438
187
196.9
86
91.6
8500 2591
185
196.0
85
91.1
BREW AT TEMPERATURE DEFINITIONS
DEFAULT: BREW AT TEMP: “ON”
(FACTORY DEFAULT FOR BREWER)
“BREW at TEMP:
-Batch will not start if tank temperature
is below set point.
-Display will show “HEATING”
and hot water tank temperature
The “BREW START” entry buttons will not
illuminate until the hot water tank reaches the
selected temperature.
Controls allow both sides of dual brewer to
operate if one side has an ongoing brew
started and the second side brew is selected.
Notifications shown on screen:
TEXT:
HEATING
Tank above 87°C/189°F-will
allow brew at low temperature.
Coffee flavor may be affected
TEXT:
L. HEAT
Tank above 76°C/169°F-will
allow brew at low temperature.
Coffee flavor will be noticeably affected
Hot water tank not
at brew temp
setpoint.
Tank temp
STOP is not lit
BREW START
buttons not lit.
and are disabled.
When hot water
tank temperature is
at setpoint. Buttons
will illuminate and
“READY” will be
displayed
vv
vv
vv
HEATING
160°F
USER SELECTABLE OPTION: BREW AT TEMP:
OFF
(Not recommended)
Unit will operate at cold temperature
Allows brewing at any temperature above 77°C/170°F
Below 70°C/170°F The red colored ready lights switch turn off
STOP
STOP