![Ferno ITC-HL User And Maintenance Manual Download Page 11](http://html1.mh-extra.com/html/ferno/itc-hl/itc-hl_user-and-maintenance-manual_554426011.webp)
11
© Ferno Washington Italia. Rel.010916
ITC - Retaining system for incubator
Incubator Installation
The incubator is anchored to the ITC-HL frame in accordance
with the manufacturer's guidelines, using the 4 anchorages
provided on the ITC-HL frame, which have the following
centre distances, 218 mm and approx. 525 mm (the latter
centre distance may be slightly changed by moving the two
bars that support the incubator, ensuring that the incubator
weight is correctly balanced on the ITC-HL frame). The
incubator frame must be provided with holes with diameter
equal to 13 mm, with the same centre distance of the
anchorages mentioned above.
If it is necessary to adapt the centre distance of the anchorages
of the ITC-HL frame with those of the incubators, slightly
loosen the 12 screws of the clamps holding the two brackets
supporting the incubator onto the ITC-HL frame, move the
two brackets symmetrically to balance the weight and fully
tighten the screws on each bracket.
Figure 13: Installation of ITC-HL frame on Ferno 5126 and 5126EL Stretcher.
INSTALLATION OF ITC-HL FRAME
ON FERNO 5126 / 5126EL STRETCHER
Warning:
to assemble the ITC-HL system on 5126 series stretchers, make a slot in the ABS central panel with a cutter or small
saw in order to fix the relevant blocks onto the frame, as shown by the two RED arrows in figure 13.
WARNING
Always apply threadlocker when indicated in this
user manual.
________________________________________
Follow the installation instructions contained in
this user manual, supplied with the product, to
guarantee that the safety requirements are met.
The declaration of conformity of the locking system
released by Ferno Washington Italia concerns the
product installed in compliance with the correct
installation instructions; Ferno Washing Italia will
not be held responsible for an incorrect installation.