5.4 Testing the Gate Antenna
After tuning the gate antenna, you can check for proper function using a Reader, the ISOStart
service software and a Transponder. Here the Noise Level and performance of the gate are tested.
5.4.1 Checking the Noise Level
Step
Action
Note
1
Select „Function Unit
Command“ „Multiplexer“
„Channel Select“ activate
antenna 1.
„Cascade Level = 1“
„Output Channel of Input 1
= 1“
2
Set by clicking on “Send”
3
Activate “Test and
Measurement”
4
Select „Noise Level“ and
start by clicking on „Start“
5
Normal Noise Level values
at antenna 1:
Average:
< 30mV
Difference (Max-Min):
< 20mV