holder. Wait until the red light on the Compustage is off before proceeding. 
(b) Remove the objective aperture by sliding the lever to the right.  
Align the specimen holder with the small pin at approximately 3 o’clock.  Carefully insert 
the end of the specimen holder into the stage until you reach the end. Then rotate 
clockwise to about 5 o’clock whilst very gently pushing inwards. This will allow the 
holder to slide in to the final pumping position.  The red light on the Compustage will 
turn on and the turbo pump will begin speeding up (orange button color) and in ~3 
minutes reach its operating speed (yellow button color). A countdown will begin in the 

Vacuum Overview

 window. Under the 

Messages/Log

 menu select the type of specimen 

holder inserted. 
When the countdown has finished and the red Compustage light has switched off, rotate 
the specimen about 120 degrees anti-clockwise (as far as it will go), then slowly slide it 
into the microscope.  Do not let go of the holder as the vacuum will pull it in and this may 
damage the stage. Once the holder is inserted click on 

Turbo

 to switch off the turbo. 

  

Cryo sample 

Under the 

Search

 tab, in the 

Stage

2

 menu, click on the flap-out menu and click 

Control

 

then 

Holder

. This will reset X, Y, Z, and 

α

 on the stage and avoid damaging the holder. 

Wait until the red light on the Compustage is off before proceeding. 
Remove the objective aperture by sliding the lever to the right.  
Under 

Setup

 tab click on 

Turbo

 to switch on the turbo pump. The turbo pump will start 

speeding up (orange button color) and in ~3 minutes reach its operating speed (yellow 
button color). Under 

Search

>

Stage

2

>

alpha toggle

. Enter “-55” then click 

Set Alpha

 to 

tilt the stage. Under the 

Setup

>

Vacuum

>

Cryo

 click on 

Prepump Airlock

. Insert the 

cryo-holder as above (adjusting the insertion position for the tilted stage).  The turbo will 
start pumping on the airlock within ~4 seconds of the Compustage red light turning on. 
The countdown will begin in the 

Vacuum Overview

 window. Under the 

Messages/Log

 

menu select 

“ST Cryo Holder”

When the countdown has finished and the red Compustage light has switched off, rotate 
the specimen about 120 degrees anti-clockwise as far as it will go, and slowly slide it into 
the microscope. Click on 

Set Alpha

 to reset the stage. Click on Turbo to switch off the 

turbo and top-up the liquid nitrogen in the holder. 
 
6. 

Removing the specimen holder

 

Reset the Stage (

Search

>

Stage

2

>

Reset Holder

).Close the column 

valves(

Setup

>

Vacuum

>

Column Valves

). Stand up and position yourself opposite the 

compustage. Use your left-hand fingers to push gently against the purple plate of the stage 
and use your right hand to slowly retract the holder outwards. When you reach a stop, 
rotate the holder clockwise ~120° until it stops. Then pull out gently and parallel to the 
stage axis. You will have to overcome the vacuum pull but do not pull too strongly as you 
may damage the stage. 

 

 

Summary of Contents for Tecnai G2 F20

Page 1: ...the ones below and highlighted in green Gun will always be 1 Log 4 Fill the cold trap dewar with LN2 Cover the screen window first Fill the dewar until it is full If you are the first user of the day...

Page 2: ...ding Remove the objective aperture by sliding the lever to the right Under Setup tab click on Turbo to switch on the turbo pump The turbo pump will start speeding up orange button color and in 3 minut...

Page 3: ...ol panel 2 Centering the condenser aperture Condenser a Select a C2 aperture 100um for TEM Single particle b At a magnification of 10 000 go to crossover and center the beam using the left trackball c...

Page 4: ...e phosphor screen using Multifunction knobs X and Y b Beam Tilt Pivot Points X and Y Click on Beam Tilt Pivot Points X and go to crossover Try to minimize the movement of the beam using Multifunction...

Page 5: ...sert the objective aperture in slide left and using the aperture position knobs on the aperture center it around the center of the un diffracted beam 6 Objective lens stigmation To stigmate the object...

Page 6: ...k Search or Preview to view the image and click Focus and move the red box to a contrasty area to adjust focus using the objective focus knob i Click Acquire to obtain an image In the TIA analysis mod...

Page 7: ...the 50K_recommended lds file or set up your low dose as below At 50 000x the pixel size is 2 08 Search Focus Exposure 2 sec integration time Magnification Detector 1 700x Orius 150 000x Orius 50 000x...

Page 8: ...Appendix Microscope Information Line resolution nm 0 144 Cs objective mm 2 0 Cc objective mm 2 0 Focal length mm 2 7 Detector FEI Falcon II Physical pixel size m 14...

Reviews: