27
or the user of this device should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment –
guidance
RF emissions CISPR 11
Group 1
This device use RF energy
only for its internal function.
Therefore, its RF emissions
are very low and are not likely
to cause any interference in
nearby electronic equipment.
RF emissions CISPR11
Class B
This device is suitable for used
in domestic establishment and
in establishment directly
connected to a low voltage
power supply network which
supplies buildings used for
domestic purposes.
Harmonic emissions IEC
61000-3-2
Class A
Voltage fluctuations / flicker
emissions IEC 61000-3-3
Complies
Guidance and Manufacturer’s declaration – electromagnetic immunity
This device is intended for use in the electromagnetic environment specified below. The customer
or the user of this device should assure that they are used in such an environment.
Immunity test
IEC 60601
test level
Compliance
Electromagnetic
environment –
guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
±8kV contact
±2kV, ±4kV, ±8kV,
±15kV air
±8kV contact
±2kV, ±4kV, ±8kV,
±15kV air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material, the
relative humidity
should be at least
30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 1kV
1 00 kHz repetition
frequency
± 1kV
1 00 kHz repetition
frequency
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surges Line-to-ground
IEC
61000-4-5
± 2kV
± 2kV
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips,
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of this
device requires
continued operation
during power mains
interruptions, it is
short interruptions
<5 %
U
T
(>95
<5 %
U
T
(>95
and voltage
% dip in
U
T
.) for
% dip in
U
T
.) for
variations
0.5 cycle
0.5 cycle
IEC 61000-4-11
<5 %
U
T
(>95 % dip in <5 %
U
T
(>95 % dip in