Recommended separation distances between   

portable and mobile RF communications equipment and the 

 

[EQUIPMENT or SYSTEM]

The   

[EQUIPMENT  or  SYSTEM]  i

s

  intended  for  use  in  an  electromagnetic  environment  in  which  radiated  RF 

disturbances  are  controlled.  The  customer  or  the  user  of  the   

[EQUIPMENT  or  SYSTEM] 

can  help  prevent 

electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications 
equipment  (transmitters)  and  the 

 

[EQUIPMENT  or  SYSTEM] 

as  recommended  below,  according  to  the  maximum 

output power of the communications equipment. 

Rated maximum 

output power of 

transmitter 

(W) 

Separation distance according to frequency of transmitter 

(m) 

 

150 KHz to 80 MHz 

P

d

167

.

1

=

 

80 MHz to 800 MHz 

P

d

167

.

1

=

 

800 MHz to 2.5 GHz 

P

d

333

.

2

=

 

0.01 

0.117 

0.117 

0.233 

0.1 

0.369 

0.369 

0.738 

1.167 

1.167 

2.333 

10 

3.689 

3.689 

7.379 

100 

11.667 

11.667 

23.333 

For  transmitters rated at a maximum  output power not listed above, the recommended separation distance d in 
metres  (m)  can  be  estimated  using  the  equation  applicable  to  the  frequency  of  the  transmitter,  where  P  is  the 
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. 
NOTE 1     At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
NOTE 2     These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption 
and reflection from structures, objects and people. 

 

 

 

 

 

 

FCC ID: 2ADNQFTC11BBT 

This device complies with Part 15 of the FCC Rules. Operation is subject to the   

Following two conditions: 

(1) This device may not cause harmful interference, and (2) This device must accept any 

interference received ,including interference that may cause undesired   

operation 

 

 

 

 

 

 

 

Summary of Contents for FT-C21Y

Page 1: ...the manual appropriately for future reference This User Manual is suitable to the following models FT C21Y FT C22Y FT C23Y FT C24Y FTC11B FT C12B FT C21Y V FT C22Y V FTC23Y V FT C24Y V FT C11B V FT C...

Page 2: ...ual by using a non invasive technique in which an inflatable cuff is wrapped around the arm The cuff circumference is limited to 8 7 inches to 12 6 inches Specification Product Name Arm Automatic Bloo...

Page 3: ...by internal electromagnetic valve Display LCD digital display It can show Pressure Pulse Date Time Color Backlight display on LCD Optional White backlight display when power on Green backlight display...

Page 4: ...o protect natural environment by respecting national and or local recycling regulations when disposing of the battery and the product at the end of their useful live II PRECAUTION FOR USE AND MAINTENA...

Page 5: ...DO NOT wrap the cuff on the arm on the side of a mastectomy 13 DO NOT use the device in the vicinity of flammable gases such as those used for anesthesia Doing so could ignite the gases and cause an e...

Page 6: ...DO NOT attempt to disassemble the device User can open battery cover for new battery installation DO NOT bend the cuff or air hose excessively 5 NEVER clean the blood pressure monitor with thinners or...

Page 7: ...n Deflation Indicator 8 Blood Pressure Measurement Unit 9 Battery Symbol 10 Irregular heartbeat Indicator 11 Memory Record Number Back light display function is optional IV Measure Procedure Battery L...

Page 8: ...ced Figure 3 Figure 4 NOTE AC adapter is optional for power supply Only use the adapter provided by manufacturer that is complying with EC60601 1 standard requirements Do not touch the AC adapter with...

Page 9: ...or Kpa and the mmHg is the definition unit When the machine is turned off press the button ON OFF more than about 10 seconds till LCD blinks then press Memory button to switch between mmHg and Kpa Arm...

Page 10: ...S CUFF PRESSURE MAY EFFECT BLOOD FLOW AND CAUSE HARMFUL INJURY Figure 6 Note The cuff is TYPE BF APPLIED PART Measuring Process POSTURE FOR TAKING BLOOD PRESSURE MONITOR Cuff should be at the same hei...

Page 11: ...will speak out the displayed blood pressure heart rate If an irregular heartbeat is detected the IRREGULAR HEARTBEAT symbol will appear and blink in the display screen See Figure9 NOTE Do not self di...

Page 12: ...This model has 2 Person memory banks and 90 memories storage for each Press and release SET button can prompt to P1 Person 1 or P2 Person 2 as your ID to access the measurement for the first time oper...

Page 13: ...85 89 Stage 1 Hypertension 140 159 90 99 YELLOW Stage 2 Hypertension 160 179 100 109 ORANGE Stage 3 Hypertension 180 110 RED Note The graph is not exact but may be used as a guide in understanding non...

Page 14: ...each measurement finish The device will speak out the last time displayed memory blood pressure systolic and diastolic pressure and heart rate when reading memory result Note The voice function is ONL...

Page 15: ...tion method After testing the result displayed on LCD including systolic pressure diastolic pressure and pulse will be automatically sent to the blue tooth receiving device such as mobile phone The ad...

Page 16: ...during the day and will also be affected by emotional and physical condition LCD shows Er U Insufficient inflation Measure again LCD shows Er H Inflation over 305 mmHg Measure again LCD shows Er 1 Un...

Page 17: ...eception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receivi...

Page 18: ...ns power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode Not applicable Mains power quality should be that of a typic...

Page 19: ...urvey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Radiated RF IEC 61000 4 3 3 V m 80 MHz t...

Page 20: ...117 0 117 0 233 0 1 0 369 0 369 0 738 1 1 167 1 167 2 333 10 3 689 3 689 7 379 100 11 667 11 667 23 333 For transmitters rated at a maximum output power not listed above the recommended separation di...

Page 21: ...PART Symbol for Follow operating instructions IP21 Symbol for the IP classification Symbol for RF transmitters Manufacturer FUDAKANG INDUSTRIAL CO LTD Address No 8 Yinghe Road Yuanjiangyuan Managemen...